Alpha MED-A64HE1 Product Manual Download Page 6

4

MED64 Head Amplifier Components and Functions

The MED64 Head Amplifier (64-CHANNEL HEAD AMPLIFIER, MED-A64HE1) is the head amplifier for the MED64 
system. Filed potentials (extracellular signals) are acquired from the acute or cultured biological preparations 
placed onto the 64 planar microelectrodes in the MED probe. The acquired raw signals are amplified by x10 with 
the MED64 Head Amplifier, then sent to the MED64 Main Amplifier (64-CHANNEL MAIN AMPLIFIER, MED-
A64MD1), where input signals are amplified further and digitized. Stimulus current can be delivered via any of 
the 64 electrodes using the built-in stimulator in the MED64 Head Amplifier.

Components and their functions

Front panel

(1) POWER
     Turns amplifier ON and OFF.

(2)STIMULUS CURRENT 
    Selection of x2 doubles the output stimulus current. Selecting NORMAL maintains the output stimulus current 

amplitude equal to the value set in the software. (e.g. when 10 

A is set in the software with x2 selected, the 

output stimulus current amplitude is 20 

A. When 10 

A is set in the software with NORMAL selected, the 

output stimulus current amplitude is 10 

A) 

    

Back panel

(1) INPUT

Signal input. Connects to the 68 pin terminal of the MED connector.

MED probe              MED connector                  MED64 Head Amp.                MED64 Main Amp.       PC and Mobius software

Stimulator

Digitizer

(1)

(2)

 

(1)

(2)

(3)      (4)         (5)          (6)

(7)

Summary of Contents for MED-A64HE1

Page 1: ...Product Manual MED64 Head Amplifier 64 CHANNEL HEAD AMPLIFIER P N MED A64HE1 A low noise multi electrode array system for in vitro electrophysiology...

Page 2: ...mages resulting from the use of information contained in this document or from the use of programs and source code that may accompany it In no event shall the publisher and the author be liable for an...

Page 3: ...e MED64 Main Amplifier to other Components 6 Signal Acquisition with the MED64 system 7 Amplifier Input Impedance Requirements for avoiding Signal Attenuation 7 Stimulation Using Planar Microelectrode...

Page 4: ...ower cords with cuts scratches or poor connectors as this may result in fire or electric shock Excessive bending pulling or slicing of the cord should also be avoided Do not pull on the cord when you...

Page 5: ...symbols used on equipment and documentation Environmental Conditions Indoor use Altitude up to 2000 m Temperature 5 40 C Maximum relative humidity 80 for temperatures up to 31 C decreasing linearly to...

Page 6: ...livered via any of the 64 electrodes using the built in stimulator in the MED64 Head Amplifier Components and their functions Front panel 1 POWER Turns amplifier ON and OFF 2 STIMULUS CURRENT Selectio...

Page 7: ...2 STIMULUS OUTPUT on the 64 CHANNEL MAIN AMPLIFIER MED A64MD1 5 CONTROL INPUT Input for control of stimulus channels Connects to the CONTROL OUTPUT on the 64 CHANNEL MAIN AMPLIFIER MED A64MD1 6 SIGNAL...

Page 8: ...idified incubator It is connected with a cable as long as 2m without noise or signal attenuation The MED64 system does NOT usually require a Faraday cage or vibration isolation table necessary for con...

Page 9: ...have very low impedances at 1kHz 10 k for 50 m electrodes 40 k for 20 m electrodes The impedances are small even at low frequencies as seen in Fig 2 Thus the input impedance of 100 M for the MED64 Hea...

Page 10: ...field potentials Vf change according to the output current Is as seen on the right in the Fig 4 Stimulus Artifacts and Biphasic Stimulation During stimulation signals stimulus pulses are recorded at...

Page 11: ...ositive components as seen in Fig 5 When biphasic stimulation is applied the electrode is charged and discharged quickly However when mono phasic stimulation is applied the electrodes need longer time...

Page 12: ...s 50 000 pF for 50 x 50 m electrodes 22 000 pF for 20 x 20 m electrodes allowing stimulation with current amplitudes up to 200 A for 0 2 msec duration for each phase is 0 1 msec DO NOT use stimulus cu...

Page 13: ...n maladjustment of user controls improper maintenance modifications or service by anyone other than AMS or damage attributable to acts of God Specifications Amplifier Stimulator General Power supply u...

Page 14: ...ntific Inc Saito Bio Incubator 209 7 7 15 Saito asagi Ibaraki Osaka 567 0085 Japan Phone 81 72 648 7973 FAX 81 72 648 7974 http www med64 com support med64 com Manufactured by Alpha MED Scientific Inc...

Reviews: