1.2.2
Optional Battery Operation
P/N 95BAT
The Series 9510 can be equipped so that it may be powered from a normal AC/DC source as well as an
optional built-in battery pack.The batteries are installed at the factory and are designed for applications
where the user desires to power the analyzer for short periods using the batteries. With this option, the
Series 9510 Carbon Dioxide Analyzer will operate for a period of at least 8 hours . If equipped with a
sample pump, operating time will be reduced to less than four hours of continuous use. The Series 9510
Carbon Dioxide Analyzer is equipped with a “smart charging” capability for the Nickel Metal Hydride
(NiMH) battery pack. Smart charging eliminates the risk of the battery pack being over-charged (not
available with explosion proof systems). To acheive a full-charge after the batteries have been discharged,
allow for approximately 16 hours of charging time. See c autionary note on page 12.
1.2.3 Optional Serial Communications / RS232C
P/N 9RS2
The RS-232C serial communications option is installed at the factory and is designed for applications
where enhanced serial communications is required between the analyzer and a host system. The
maximum distance between the analyzer and host system is 50 feet
1.2.4 Optional Serial Communications / RS485
P/N 9RS4
The RS-485 serial communications option is installed at the factory and is designed for applications where
enhanced serial communications is required between one or several analyzers and a host system over the
same communications channel. The RS-485 format allows both sending and receiving of signals over
greater distances than RS-232C, making it ideal for installations where the analyzer(s) are located at
distances greater than 50 feet from the host. The maximum distance recommended between devices is
4,000 feet
The following options are only available with the flow through system configuration only:
1.2.5 Optional Particulate Sample Filter
P/N 995S
Miniature T-type 316 stainless steel filter with 1/8" NPT in-line ports. Recommended when particle loading
exceeds 3 mg/ft
3
and hydrocarbon mist exceeds 0.7 mg/ft
3
. Installed at the factory unless otherwise
specified by the customer.
1.2.6
Optional Low Capacity Pressure Regulator
P/N 9LPR
Pressure Regulator with aluminum body recommended for use when sample pressure is
below 100 PSIG. The regulator does not include a gauge (can be ordered separately).
1.2.7
Optional Sample Pump
P/N 9PMP
Addition of a sample pump is designed for those applications where the sample pressure or flow is
insufficient to transport the sample through the sensor housing.
1
.2.8 Optional Coalescing Filter
P/N 9CF
Coalescing filter recommended when the sample may contain a light mist together with particles.
1.2.9
Optional Horn & Strobe
P/N Spec-HnSt
The Series 9510 can be equipped with an optional 24 VDC or 115 VAC Horn & Strobe for remote alarm
indication (separate from the Series 9510 Electronics that has its own audible and visual alarm indicators).
* Please contact the factory for other options and accessories.
ALPHA OMEGA INSTRUMENTS CORP.
SERIES 9510 CARBON DIOXIDE ANALYZER INSTRUCTION MANUAL