SECTION 3.0
OPERATION
3.1 Preparation for Operation
3.1.1 AC Input Voltage Selection
The Series 9510 Carbon Dioxide Analyzer has an ON/OFF power switch on the side of the instrument.
The instrument is shipped to operate on the line voltage specified in the purchase order. If the line voltage
is not specified, the instrument will be configured for operation at 115 VAC, 50-60Hz. If the instrument is
going to be powered from an external DC power source, please refer to section 2.2.1 . Please note that if
an option such as a sample pump has been ordered, the power to instrument must be in accordance with
the power requirements of the option in question. For instance, you cannot power the Series 9510 with
115 VAC if it has been equipped with a 24 VDC sample pump. Please check the original purchase order.
The input voltage can be changed in the field by changing the S02 jumpers located on the lower left
quadrant of the printed circuit board (refer to FIGURE 1.0). The Series 9510 Carbon Dioxide Analyzer is
configured for operation with a .5 ampere slo-blow fuse. The Fuse type is Wickman WK4041BK.
Changing the AC input voltage can be done as follows:
1. Turn Power off and disconnect the Series 9510 Carbon Dioxide Analyzer from all AC power.
2. Swing out the front panel to gain access to the main printed circuit board
(refer to Section 2.1 for instructions).
3. Remove the cover on the lower left of the main circuit board, and examine the jumpers on the main
circuit board. Shown below are the two ways of installing the AC input selection jumpers at S02.
2 3 0
1 1 5
S02
S02
4. To change the AC configuration, remove the jumpers and place them according to the above
diagram. Be sure to replace the cover.
5. Reassemble the instrument and proceed to the next section.
3.1.2 Mounting Configurations
Bench Top / Portable (BTP Configuration)
The BTP is the basic configuration for the Series 9510
Carbon Dioxide Analyzer, designed for either bench top or portable applications. All gas and electrical
connections are located on the instrument's rear panel for ease of access.
Panel Mounting (PNL)
- In the PNL configuration, a panel mounting frame is supplied. Make the
appropriate panel modifications as indicated in the drawing in Appendix B.
Explosion Proof Enclosure
(EXX).
Please refer to the Appendix C