20
or equal to the Filter Threshold value, N1, then Filtering is ACTIVE. If this
condition occurs for N2 consecutive A/D readings the filtering is
DE-ACTIVATED
and the Current Value is set equal to the New A/D value as follows:
IF: AD -
AD’
≤
$G,
THEN:
Filtering is ACTIVE
IF: AD -
AD’
> $G for N consecutive readings
THEN:
Filtering is NOT ACTIVE
Where:
(AD) is the
new
A/D reading for the Device
(AD’) is the
previous
A/D reading for the Device
($G) is the averaging threshold
value
(N) is the averaging threshold
count
Note: See Legend for Device (D) location mapping
Response:
H
(D) =
[current value > new value]
<CR><LF>
$I
–
INITIALIZE DEVICE CALIBRATION
Syntax:
$I
(D)
Where:
(D) is the device (channel) (0-7)
Initializes the calibration kept in the Device Parameter Array (DPA) for the
specified Device as follows:
GAIN =
1
OFFSET = 0
If a Calibration File
IS
being used, deleting all the calibration data but leaving the
file intact initializes the calibration file.
Response:
Device
(D)
Initialized
<CR><LF>