Allora Positive Contact PCS-250 Installation Instructions Manual Download Page 9

Page 8

PCS-250 System

Trouble Shooting Guide

NO POWER TO TCU-250

(No lights illuminated when TCU-250 is powered up and no power to PCS-250.)

If 115VAC, is the power supply supplied by Allora being used?

Is there a clean 24VDC being supplied to the power terminals on the TCU-250?

Is there no more than two TCU-250's running on one power supply?

Is there 110-120VAC being supplied to the power supply?

Is the power supply sending 24VDC to the power terminals on the TCU-250?

NO POWER TO PCS-250

Are one or more lights on the TCU-250 illuminated?

If not, see 

No Power to TCU-250

Is the PCS-250 connected properly?

Color coded wires of cable should be attached tot he correct terminals on the TCU-250.

Cable should be screwed tight to the PCS-250 connector.

The inside of the PCS-250 connector should be dry.

Manually move the sensing needle away from the start position.  Does the sensing needle return to the start 
position on its own?

Is there approximately 4.1VDC across the blue and black terminals on the TCU-250?  (The PCS-250 
must remain connected to TCU-250 for this measurement.)

Is there 24VDC across the Power terminals on the TCU-250?

If not, see 

No Power to TCU-250

NO FAULT SIGNAL WHEN TOOL IS BROKEN OR FAULT SIGNAL WHEN TOOL IS OKAY

Was the TCU-250 taught the tool's position?

To teach the TCU-250, press and release the teach button. The yellow teach light will illuminate. 
Press and release the teach button a second time, or give the TCU-250 a start signal. The TCU-250 
will run through one cycle, the yellow light will turn off, and the green light will illuminate.
The TCU-250 is now taught that tool's current position.

Is the tool within the PCS-250's swing angle?

The sensing needle's tip should extend over the tool's centerline

The sensing needle should contact the tool at least 1.60mm from the tool's tip.

Is the sensing needle supplied by Allora being used?

Is the clamping collar tight on the PCS-250?

Is the swing arm assembly tight on the PCS-250?

Is there enough time set on the TCU-250 for the PCS-250 to reach its full swing angle?

Have any objects obstructing the swing angle of the PCS-250 been removed?

With power disconnected from PCS-250, can the sensing needle be moved freely back and forth without any 
binding?

Look for metal chips which may be packed around the seals on the PCS-250 or between the PCS-250 
and the swing arm assembly.

Look for large chip "bird's nests" that may be getting caught on the PCS-250, or on the tool.

Is the PCS-250 connected properly?

Color coded wires of cable should be attached to the correct terminals on the TCU-250.

Cable should be screwed tight to the PCS-250 connector.

The inside of the PCS-250 connector should be dry.

Is the PCS-250 returning to its start position at the end of each cycle?

The V-seal in the swing arm assembly provides additional protection against contaminants and should not be 
removed in most applications.  The V-seal may however restrict the sensors motion when the swing rate is set 
on “LO”. If your application does not allow “HI” rate operation, you may remove the V-seal from the swing 
arm assembly to determine if this solves the problem.

Summary of Contents for Positive Contact PCS-250

Page 1: ...0 OK FAULT TEACH POWER 24VDC CW LO NT TIME FAULT OUTPUT OK OUTPUT COM 0 BLUE BLACK BROWN WHITE GRAY 1 2 3 4 6 5 4 3 2 1 CCW 1 2 3 HI NO NO NC NC M M L L TEACH RESET START 115V 24V ET USA Made in the P...

Page 2: ...raight connector available upon request PCS 250 SENSOR 40MM 1 57 10MM 0 393 140MM 5 50 120MM 4 72 28MM 1 102 37 5MM 1 475 27MM 1 062 5 0M 16 0 TCU 250 OK FAULT POWER 24VDC CW LO NL TIME FAULT OUTPUT O...

Page 3: ...ace available in the tooling area will determine the easiest mounting method and location With all machine power off hand hold the sensor in the tooling area to imagine its best placement Remember the...

Page 4: ...to be repositioned while attached to the sensor rotate the entire sensor to avoid damage to the connector Make sure that the dip switch on the control unit is set to correspond to the CW or CCW direc...

Page 5: ...application T 16 is the common for either a 115VAC start signal T 14 or a 24VDC start signal T 15 The TCU recognizes a change of state in the circuit to activate the checking sequence A minimum durat...

Page 6: ...ount of time the sensor rotates during the scanning sequence Since it takes a greater amount of time to swing the sensing needle 180 than 90 a greater amount of time needs to be set If the sensor will...

Page 7: ...d 115VAC machine control voltages For any questions or installation assistance contact 1 855 866 5911 TCU 250 OK FAULT TEACH POWER 24VDC CW LO NL TIME FAULT OUTPUT OK OUTPUT COM 0 BLUE BLACK BROWN WHI...

Page 8: ...gnal Terminals 14 16 are for a 115VAC start signal Is the correct start signal being applied The signal can be AC or DC voltage Polarity does not matter The TCU 250 can be started with a low high low...

Page 9: ...l The TCU 250 will run through one cycle the yellow light will turn off and the green light will illuminate The TCU 250 is now taught that tool s current position Is the tool within the PCS 250 s swin...

Page 10: ...d mounting bracket for the Filter Regulator Gage is included with AP 1000 Kit Connect the provided NPT fitting into the filter regulator gage in the normal fashion Cut the tube to the required length...

Page 11: ...ingle AP Sensors are supplied with an air tube and a NPT fitting for connection to the filter regulator Some machines may utilize more than one sensor One Allora AP 1000 Kit or customer supplied equiv...

Page 12: ...ompliant materials They are free of Bromide Halogens PBB PBDE Mercury Cadmium and Chrome 6 All Positive Contact products are Lead Free All Positive Contact products comply with current EU environmenta...

Reviews: