Allora Positive Contact PCS-250 Installation Instructions Manual Download Page 4

1.60mm

4)

Mounting the bracket and sensor

Mount the fabricated bracket and attach the clamping ring. Slide the sensor 
into the clamping ring but do not tighten in to place. Slip on the swing arm 
assembly ( you do not need to secure it ) and move the sensor forward or 
backward until the sensing needle is positioned approximately 1.60 mm back 
from the tip of the tool as shown in FIG 4. Also make sure that the sensing 
needle extends past the center line of the tool. When this position is realized, 
secure the sensor body into place by tightening the clamping ring. Remove 
the swing arm assembly.

FIG 4

5)

Final sensor set-up

Install the sensor control unit into the machines electrical cabinet in accordance 
with the TCU installation instructions. Connect the cable to the sensor and power 
up the control unit. Be sure the cable is properly positioned to avoid any 
interference with the machines operation. NOTE: if the cable needs to be 
repositioned while attached to the sensor, rotate the entire sensor to avoid 
damage to the connector. Make sure that the dip switch on the control unit is set 
to correspond to the CW or CCW direction you have selected. Put the swing arm 
assembly back on and position it where you want the needle swing to start.

Note: The most common swing angle is 90º or less, but you may start up to 300º 
away from the tool if necessary for needed clearance.

Once you have selected the starting point, secure the swing arm assembly to the 
sensor. The sensor is now ready to learn the location of the tool. Press the teach 
button on the control unit. The yellow LED will illuminate indicating the system is 
ready to learn the tool position. Press the teach button again or apply a start 
signal. The needle will swing and contact the tool. The sensor has now learned 
the location of the tool. Once another start signal is applied and the sensor’s 
swing angle is not repeated ( plus or minus a small factory preset value) a fault 
condition will occur. Refer to the Teach Button section in the TCU installation 
instructions for more details.

Page 3

Summary of Contents for Positive Contact PCS-250

Page 1: ...0 OK FAULT TEACH POWER 24VDC CW LO NT TIME FAULT OUTPUT OK OUTPUT COM 0 BLUE BLACK BROWN WHITE GRAY 1 2 3 4 6 5 4 3 2 1 CCW 1 2 3 HI NO NO NC NC M M L L TEACH RESET START 115V 24V ET USA Made in the P...

Page 2: ...raight connector available upon request PCS 250 SENSOR 40MM 1 57 10MM 0 393 140MM 5 50 120MM 4 72 28MM 1 102 37 5MM 1 475 27MM 1 062 5 0M 16 0 TCU 250 OK FAULT POWER 24VDC CW LO NL TIME FAULT OUTPUT O...

Page 3: ...ace available in the tooling area will determine the easiest mounting method and location With all machine power off hand hold the sensor in the tooling area to imagine its best placement Remember the...

Page 4: ...to be repositioned while attached to the sensor rotate the entire sensor to avoid damage to the connector Make sure that the dip switch on the control unit is set to correspond to the CW or CCW direc...

Page 5: ...application T 16 is the common for either a 115VAC start signal T 14 or a 24VDC start signal T 15 The TCU recognizes a change of state in the circuit to activate the checking sequence A minimum durat...

Page 6: ...ount of time the sensor rotates during the scanning sequence Since it takes a greater amount of time to swing the sensing needle 180 than 90 a greater amount of time needs to be set If the sensor will...

Page 7: ...d 115VAC machine control voltages For any questions or installation assistance contact 1 855 866 5911 TCU 250 OK FAULT TEACH POWER 24VDC CW LO NL TIME FAULT OUTPUT OK OUTPUT COM 0 BLUE BLACK BROWN WHI...

Page 8: ...gnal Terminals 14 16 are for a 115VAC start signal Is the correct start signal being applied The signal can be AC or DC voltage Polarity does not matter The TCU 250 can be started with a low high low...

Page 9: ...l The TCU 250 will run through one cycle the yellow light will turn off and the green light will illuminate The TCU 250 is now taught that tool s current position Is the tool within the PCS 250 s swin...

Page 10: ...d mounting bracket for the Filter Regulator Gage is included with AP 1000 Kit Connect the provided NPT fitting into the filter regulator gage in the normal fashion Cut the tube to the required length...

Page 11: ...ingle AP Sensors are supplied with an air tube and a NPT fitting for connection to the filter regulator Some machines may utilize more than one sensor One Allora AP 1000 Kit or customer supplied equiv...

Page 12: ...ompliant materials They are free of Bromide Halogens PBB PBDE Mercury Cadmium and Chrome 6 All Positive Contact products are Lead Free All Positive Contact products comply with current EU environmenta...

Reviews: