Command reference
R&S
®
ZND
941
User Manual 1173.9557.02 ─ 61
Note
that during the corresponding calibration sweep the R&S
ZND expects the physi-
cal port connections to be established exactly as specified by the created port assign-
ments.
Use
●
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment:DEFine:TPORt:
DEFault
on page 942 to use auto-detection instead of defining "test port --> cal
unit port"-connections explicitely
●
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment<Asg>:DEFine
or
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment<Asg>:DEFine:
TPORt
to define (additional) port assignments manually
See
Chapter 4.5.5.6, "Multiple port assignments"
on page 173 for background infor-
mation.
Suffix:
<Ch>
.
Channel number
Setting parameters:
<TestPort1>
Test port number 1
<TestPort2>
Test port number 2
Example:
[SENSe<Ch>:]CORRection:COLLect:AUTO:
Usage:
Setting only
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment<Asg>:DEFine:TPORt
<TestPort1>[, <TestPort2>]...
Similar logic as
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment<Asg>:
, but initially defines the port assignment only by its underlying test port set; the
connected calibration unit ports are
auto-detected
at the start of the corresponding
calibration sweep.
Suffix:
<Ch>
.
Number of the channel
<Asg>
Number of the port assignment
Parameters:
<TestPort1>
First test port number
<TestPort2>
Second test port number ...
Example:
[SENSe<Ch>:]CORRection:COLLect:AUTO:
SCPI command reference