Command reference
R&S
®
ZND
628
User Manual 1173.9557.02 ─ 61
7.3.1.2
CALCulate:DATA...
The
CALCulate:DATA...
commands provide access to the results of a measure-
ment.
Data format
The trace data is transferred in either ASCII or block data (REAL) format, depending
on the
setting. If block data format is used, it is recommended to
select EOI as a receive terminator (
SYSTem:COMMunicate:GPIB[:SELF]:
EOI
).
...........................................................................................631
CALCulate<Ch>:DATA:CALL:CATalog?
............................................................................632
CALCulate<Ch>:DATA:CHANnel:ALL?
.............................................................................632
CALCulate<Ch>:DATA:CHANnel:DALL?
.......................................................................... 633
.................................................................................................633
CALCulate<Ch>:DATA:MDATa:INTerpolate
.......................................................................634
CALCulate<Chn>:DATA:NSWeep:COUNt?
.......................................................................634
CALCulate<Chn>:DATA:NSWeep:FIRSt?
......................................................................... 634
CALCulate<Chn>:DATA:NSWeep[:LAST]?
....................................................................... 635
.................................................................................... 636
................................................................................. 636
................................................................................................ 636
CALCulate<Chn>:DATA
<Format>, <Data>...
CALCulate<Chn>:DATA?
<Format>
The query reads the response values of the selected channel's active trace or reads
error terms of the selected channel.
The "set command" either imports formatted or unformatted trace data to the selected
channel's active trace (see
on page 82) or writes error
terms of the selected channel.
Note
●
The data format is parameter-dependent; see tables below. The unit is the default
unit of the measured parameter; see
CALCulate<Ch>:PARameter:SDEFine
●
Unformatted trace data (SDATa) can only be imported to memory traces
●
Formatted trace data (FDATa) can only be imported to "live" traces if the related
channel is in single sweep mode (
OFF
).
Before the import, the target trace must be prepared according to the settings used
during export. Any request for new data from the hardware ("Restart Sweep" in sin-
gle sweep mode or switching to continuous sweep mode) discards imported data
and switches back the display to measured data.
Suffix:
<Chn>
.
Channel number used to identify the active trace
SCPI command reference
www.allice.de
Allice Messtechnik GmbH