Command reference
R&S
®
ZND
1052
User Manual 1173.9557.02 ─ 61
Example:
Starting with a 4-port analyzer's default logical port assignment
L1<-->P1, L2<-->P2, L3<-->P3, L4<-->P4,
the remote command
SOUR:LPOR1 1,2
by default (
ALIGn=ON
) generates aligned logical ports
L1<-->P1&P2, L2<-->P3, L3<-->P4.
With
ALIGn=OFF
, the result is
L1<-->P1&P2, L3<-->P3, L4<-->P4
and L2 does not exist.
Manual operation:
See
SYSTem:COMMunicate:GPIB[:SELF]:RTERminator
<Terminator>
Sets the receive terminator of the analyzer. The receive terminator indicates the end of
a command or a data block.
The receive terminator setting is relevant if block data is transferred to the analyzer
(
REAL
). In the default setting LFEoi, the analyzer recognizes an LF
character sequence with or without the EOI control bus message as a receive termina-
tor. An accidental LF in a data block can be recognized as a terminator and cause an
interruption of the data transfer.
The EOI setting is especially important if commands are transferred in block data for-
mat, because it ensures that the parser for command decoding is activated by the ter-
minator only after the command has been completely transferred. Readout of binary
data does not require a change of the receive terminator.
Note:
since firmware V2.80 this command raises an error if no GPIB interface (option
R&S
ZND-B10) is available.
Parameters:
<Terminator>
LFEoi | EOI
LFEoi
A line feed character sequence with or without EOI is recognized
as receive terminator
EOI
Only EOI is recognized as receive terminator
Example:
SYST:COMM:GPIB:RTER EOI
Set the terminator to EOI.
SYSTem:COMMunicate:NET:HOSTname
<HostName>
Sets or gets the host name of the instrument.
Parameters:
<HostName>
Host name
*RST:
A preset of the instrument doesn't change the host
name.
SCPI command reference
www.allice.de
Allice Messtechnik GmbH