SYSTem:ERRor:DISPlay[:REMote]
<Boolean>
Disables or enables the display of information popups for remote control errors. These
popups appear at the bottom of the remote screen and the manual screen.
Note
●
Display of information popups can be globally disabled/enabled using
OFF/ON
●
For SCPI error
-113, Undefined header
no tooltip is displayed
Parameters:
<Boolean>
ON | OFF - enable/disable display of information popups for
remote control errors
*RST:
ON
Example:
SYST:ERR:DISP:STATe ON; REM ON
Switch the display of information popups for remote command
errors on.
FREQ:STAR -1
Generate a
Remote Error: -222, "Data out of
range; ..."
tooltip.
Manual operation:
"Show Remote Error Info Messages"
SYSTem:ERRor:DISPlay:STATe
<Boolean>
Globally defines whether instrument events shall be indicated by information popups.
Display of popups can be limited to certain event types using commands
.
Parameters:
<Boolean>
ON | OFF - globally enable or disable the display of information
popups
*RST:
ON
Example:
and
.
Manual operation:
SYSTem:ERRor[:NEXT]?
Queries and at the same time deletes the oldest entry in the error queue. Operation is
identical to that of
The entry consists of an error number and a short description of the error. Positive
error numbers are instrument-dependent. Negative error numbers are reserved by the
SCPI standard; see section
Error messages and troubleshooting
.
SCPI command reference