Message
Action
Step Start Alarm
Follow the instruction of the message that have been created for this
message. Example: "Add blood"
Alarm 'Close Vessel Lid'
The lid is not properly closed. Release the pressure from the vessel by
using the safety while the temperature is below 80°C. Then open the lid
and close it again to get blue light coming one. If not
→
Call service.
Alarm 'Close Safety Covers'
One of the safety lids (door trap or lid to access to the dispensing port)
is not closed or not properly closed--> Close the lids
Heating error
‘
Abnormal
temperature variation
’
Time to reach the sterilization set point is too long or the temperature is
not stable --> Make sure that the stirrer is running properly. If not, check
the bearing state.
Cooling error
‘
Abnormal
temperature variation
’
Time to reach the dispensing set point is too long or the temperature is
not stable --> Make sure that the water tap is open.
Temperature warning
‘
temperature is out of tolerance
’
The temperature was not in +/-1°C range during the sterilization plateau-
Make sure that temperature is within the tolerance of your application
before validating the cycle.
Restart cycle after power on
Just for information - Check the data records at the end of the cycle
before validating the cycle.
'Media' temperature probe error
‘
Abnormal temperature
measurement
’
The vessel temperature probe is defective --> Call service
Technical Error
‘
Stirrer motor
’
Defective stirrer --> Call service
Technical Error
‘
Pressure release
solenoid
’
Defective pressure release valve --> Call service
Technical Error
‘
Cooling
’
Defective cooling valve --> Call service
Technical Error
‘
Locker magnet
’
Defective lid locking solenoid --> Call service
Communication Initialisation
failed
Switch off and on again the machine - Call service if it does not solve the
problem
Communication failed. Please
switch off and contact the after-
sales service
Switch off and on again the machine - Call service if it does not solve the
problem
•
Icônes
•
Communication between HMI and automaton cards normal
•
Failure of communication between HMI and automaton cards normal
•
SD card connected and compliant
•
SD card failure
•
Connected to an Ethernet network
- USB key connected
Summary of Contents for MEDIAWEL 10
Page 2: ...2 Sommaire de langues Fran ais 8 English 59...
Page 7: ...7 VERSION FRANCAISE...
Page 23: ...23 Modifier ou supprimer Se r f rer la modification ou suppression des programmes...
Page 27: ...27...
Page 46: ...46 ENGLISH VERSION...
Page 66: ...66...