Connection
It is essential to fit the power cable with a plug that is suitable for the above power.
Water
Inlet
•
For more convenient use, include an easy to access tap close to the MEDIAWEL, at a height of
approximately 1 meter and slightly offset at the side of the device as illustrated below
•
Cut the 4 meter hose to the desired lengths for supply and outlet hoses
•
Fit the in-line filter (ref: MEDW3008) to the water supply hose
•
Connect the supply and outlet hoses as set out below
Repère
Désignation
Référence
1
In-line water circuit filter (recommended option)
MEDW3008
2 et 3
Inlet and outlet water tubing
MEDW4012
Outlet
The cooling water drainage system must be designed to withstand a temperature of 130°C. It may be made
from copper or PVC-C or any other material that is suitable for this use.
Power supply
The adapted plug must be used adapted to the above specifications :
-
Phase
wire is identifyed by
1,
printed on one of the 2 black cables
-
Neutral
wire is identifyed by
2,
printed on the second black cable
-
Earth
wire is yellow and green
The wall power supply must be between 1 and 1,2m height and equipped with a shutdown system.
Initial commissioning
Switching on
Prior to switching the device on, you must ensure that the mains voltage between phase and neutral
complies with 230V +/- 10%. Plug the machine using the plug into the wall power supply and switch on the
front switch.
Water inlet
Water outlet
In-line water supply
(ref :
MEDW3008)