background image

29

11

1

1

1

Set the Alarm (up to 4 di

ff

erent Alarms a day)

The Alarm times 

fl

ash. Press and release SIDE(UP/DOWN button) buttons until 

you get the alarm times (0 to 4th) setting that you want. If you press and release 

the power button, the meter will advance to the Alarm Time Setting.

Set the Average Period

The Average Period 

fl

ashes. Press and release SIDE(UP/DOWN button) buttons 

until you get the period of average setting that you want (1~28d). If you press 

and release the power button, the meter will advance to the Alarm Setting.

Set the Alarm Time 

The setting for the  Alarm Time Setting will start 

fl

ashing. Press and release 

SIDE(UP/DOWN button) buttons until you get the time that you want. The Hour 

of the Alarm Time 

fl

ashes. Press and release SIDE(UP/DOWN button) buttons to 

set the alarm time the same as the Time Setting( - ). 

Save the Setting Information

Press and release the power button, and then the meter will turn o

ff

The setting information you entered will be saved.

Summary of Contents for AGM-2200

Page 1: ...ose Monitoring System 7102 7107 7402 7403 7406 140 Beolmal ro Dongan gu Anyang si Gyeonggi do 431 804 REPUBLIC OF KOREA www allmedicus com Manufactured by All Medicus Co Ltd EU Representative MT Prome...

Page 2: ...tton to turn on the meter Check the code number The code number displayed on the meter must match the one on the test strip vial Insert a test strip The meter will automatically turn on Apply the bloo...

Page 3: ...This User Manual will tell you all you need to know about the system and how it works Please read this carefully before using the meter Thank you for using Blood Glucose Monitoring System...

Page 4: ...cose testing in fresh capillary whole blood drawn from the fingertips palm arm thigh and calf The GlucoDr SuperSensor Blood Glucose Monitoring System is for testing outside the body in vitro diagnosti...

Page 5: ...between 0 40 32 104 Store the GlucoDr SuperSensor Meter at a place where relative humidity is less than 85 Keep the GlucoDr SuperSensor Meter in sanitary environment Keep the meter away from direct su...

Page 6: ...eter Caring for GlucoDr SuperSensorTM Meter Additional Information Table of Contents Entering the EVENT RECORD MODE l 21 Understanding Your Test Results l 23 Recalling the Previous Test Results l 24 C...

Page 7: ...Monitoring System 1 Carrying case User Manual GlucoDr SuperSensorTM Meter GlucoDr SuperSensorTM Test Strips Battery Lancing Device Lancets Upon request of distributor actual package may consist of al...

Page 8: ...econds CODE MODE 5 seconds SET MODE Delete all stored test results in the MEMORY MODE UP Button DOWN Button Change the code number in the CODE MODE Recall the stored test result in the MEMORY MODE Sel...

Page 9: ...Battery Cover Data Port Transfer your test results from the meter to a computer...

Page 10: ...strip to the meter BLOOD DROP ICON When the BLOOD DROP ICON blinks apply blood sample TEMPERATURE ICON When the meter is not used at a recommended temperature range Er1 andTEMPERATURE ICON appear CURR...

Page 11: ...s part to insert or remove test strip Grip Make sure the confirmation window fills completely Confirmation window Reaction chamber Insert into test port Gold electrodes Apply blood sample here blood i...

Page 12: ...on the test strip vial Checking the Code Number Press the power button to turn on the meter Then the code number STRIP ICON and BLOOD DROP ICON will appear Turn on the Meter Make sure that the code nu...

Page 13: ...tton button to adjust the same code number printed on your test strip vial Change the Code Number Check the code number on the test strip vial before inserting the test strip You can not change the co...

Page 14: ...correctly adjusted press and release the power button A new code number will be saved and appear on the meter display Save the Code Number To change the code number faster press and hold any of SIDE...

Page 15: ...meter and test strip When you leave your test strip vial cap open for a long time When you open a new vial of test strips When you think that the test results are not accurate When you drop the meter...

Page 16: ...t of the meter Gently push it all the way in until it goes no further The meter will turn on automatically and beep Then the code number and the BLOOD DROP ICON will appear Make sure that the code num...

Page 17: ...y countdown from 10 to 1 then the test result will appear Test Result Appears in 10 seconds To ensure accurate test results Gently shake the control solution vial to ensure the control solution is mix...

Page 18: ...range printed on the test strip vial is for the GlucoDr SuperSensor Control Solution only It is used to check the performance of your meter and test strips It is NOT a recommended range for your bloo...

Page 19: ...n buttons of the meter to select ATTENTION ICON Press and release the power button if selection is completed The control solution test results will be saved Record the Control Solution Test Results in...

Page 20: ...lancet it will be used to dispose the lancet safely after its use Insert a Lancet into the Lancet Holder Turn the cap of lancing device clockwise until it fits Replace the cap Before using the lancing...

Page 21: ...Charge the Lancing Device Place the lancing device carefully on the tip of your finger and press the release button Lance your Finger Place the protective disk on a flat surface With the used lancet...

Page 22: ...sterilized lancets are BD Ultra Fine E Z Ject G P Lite Microlet Monolet Soft Touch Ultra TLC Unilet GP and Greenlan Note Not recommended lancets in the above list may not work properly with the lanci...

Page 23: ...17 Performing the Blood GlucoseTest with GlucoDr SuperSensorTM GlucoDr SuperSensorTM Meter GlucoDr SuperSensorTM Test Strip Lancing Device Lancets...

Page 24: ...ber displayed on your meter and the test strip code number match Insert a Test Strip Do not force the test strip into the meter This may cause malfunction of the meter Obtain a blood drop from your fi...

Page 25: ...o the arrow on the test strip as directed Do not put blood on the top of the test strip Apply the blood drop to the side edge of the yellow window of the test strip If there is an insufficient amount...

Page 26: ...more effective diabetes management Once the meter completes the countdown from 10 to 1 your test result will appear on the meter display If you remove the test strip immediately only the test result w...

Page 27: ...Attention Press SIDE UP DOWN button buttons to select an icon which correlates with the test result Select the EVENT ICON After the selection of an EVENT ICON remove the used test strip After discharg...

Page 28: ...ol solution tests or improper blood glucose test Record all control solution tests with ATTENTION ICON to distinguish them from blood glucose tests in the meter memory Recorded control solution test r...

Page 29: ...eter and test strip are working correctly you must contact your healthcare professional immediately Expected Values1 Criteria for the diagnosis of non diabetes Fasting plasma glucose FPG 70 100 mg dL...

Page 30: ...s with the meter turned on Press SIDE Buttons 1 You can select the number of days for the average 1 28d in the SET MODE 7d shown in the left side indicates that the value 96 is a calculated average fo...

Page 31: ...p to 250 test results Recall your Test Results 25 Press and release the power button once to exit from the MEMORY MODE Exit the MEMORY MODE To delete all the stored test results in the MEMORY MODE pre...

Page 32: ...lt will help you and your healthcare professional to track changes in your blood glucose level 1 To exit the SET MODE at any time press and hold the power button for 3 seconds The settings you have al...

Page 33: ...ur If you press and release the power button the meter will save hour and advance to the Minutes Setting Set the Hour The Minutes flash Press and release SIDE UP DOWN button buttons until you get corr...

Page 34: ...eter will advance to the Day Setting Set the Month The Date flashes Press and release SIDE UP DOWN button buttons until you get the date format that you want If you press and release the power button...

Page 35: ...u get the period of average setting that you want 1 28d If you press and release the power button the meter will advance to the Alarm Setting Set the Alarm Time The setting for the Alarm Time Setting...

Page 36: ...nt alarms can be set You can continually set alarm right after the first setting Press and release the power button after you complete the first alarm time setting Then 2 appears you can repeat to set...

Page 37: ...can test a few more times but the battery should be replaced to new one as soon as possible If bAt appears on the meter display the meter will not operate The battery must be immediately replaced Alw...

Page 38: ...until it locks into place Close the Battery Cover Remove the old battery and put the new one with sign facing up Replace the Battery Press down on the battery cover then slide it out The battery cover...

Page 39: ...ult You should check that your meter matches the example exactly every time your meter turns on Do not use the meter if the display check screen does not exactly match the example Please contact your...

Page 40: ...mmol L No action is required No action is required Select an EVENT ICONs using SIDE UP DOWN button buttons before discarding the used test strip Repeat the test using a new test strip and or check the...

Page 41: ...curate test results Repeat the test using a new test strip and or check the system with the GlucoDr SuperSensor Control Solution If the test result is Lo again even though your system is working corre...

Page 42: ...used outside the proper range of temperature The used test strip was inserted in the test port No action is required No action is required Leave the meter at a temperature between 10 40 50 104 for at...

Page 43: ...was an electronical noise during the test The meter has an internal electric circuit problem The meter has a problem with strip connection Retest with a new test strip Do not remove the test strip unt...

Page 44: ...own or incompletely Replace the battery Check if battery is correctly installed with the positive sign facing up See pages 31 32 Contact your local distributor Insert the test strip correctly with the...

Page 45: ...System AGM 2200 Electrochemicalmethod Freshcapillary venous andarterialwholeblood Minimum1 5 20 900mg dL 1 1 50mmol L 10seconds Plasma equivalent One3 voltlithiumbattery coincelltypeCR2032 Approximate...

Page 46: ...cose concentration 75 mg dL 4 2 mmol L Repeatability using venous whole blood Level 42 1 mg dL 105 mg dL 142 mg dL 200 mg dL 345 mg dL 2 6 mg dL 3 2 mg dL 4 1 mg dL 4 9 mg dL 7 6 mg dL SD 6 3 3 0 2 9...

Page 47: ...Symbols 1 CAUTION This product fulfills the requirements of Directive 98 79 EC on in vitro diagnostic medical device For in vitro diagnostic use Do not dispose the instrument in the urban waste Consu...

Page 48: ......

Reviews: