Recalling the Previous Test Results
The GlucoDr SuperSensor Meter has a memory capacity to store up to 250 most recent test
results with date, time and event. It also provides you averages of your blood glucose test
results over a period of time (1~28d). You can review the test results stored in the memory
by the following steps.
To enter the MEMORY MODE, press any of SIDE(UP/DOWN button) buttons
with the meter turned on.
Press SIDE Buttons
1
You can select the number of days for the average (1~28d) in the
SET MODE. “7d” shown in the left side indicates that the value 96 is a
calculated average for the past 7 days.
Indicate the Average Value
If there is no test result, the meter will not display anything and SIDE
(UP/DOWN button) buttons will not operate.
When the whole test results' average is calculated, individual event records are
not considered. The GlucoDr SuperSensor Blood Glucose Monitoring System will use
900 mg/dL (50 mmol/L) for “HI” results, and 20 mg/dL (1.1 mmol/L) for “Lo” results.