Assigning Individual Addresses to Flow or Pressure Devices

In order to communicate with multiple units through a single RS-485 serial port, 

each 

unit must have a unique address.
On units with a standard display and manufactured after November 2005,

 the device 

address can be changed via the RS-485 Communication menu on the front panel of the 

device as outlined below.
Change the address from the front panel of a flow or pressure device as follows:

1.  Push the MODE button until the SELECT MENU screen is shown.
2.  Push the lower left button dynamically labeled “Comm. RS-485”.
3.  Push the upper left button dynamically labeled “Select” once to move the selection 

arrow down one line to “Unit ID (Current Address)…(New Address)

4.  The default address is A.
5.  Use the lower left and middle button dynamically labeled “UP” or “DOWN” to change 

the address to any letter between A and Z that is unique from other units you intend 

to connect to the Multi-Drop Box.

6.  The new address is shown at the end of the dotted line.
7.  When the desired new address is shown at the end of the dotted line, press the 

MODE button to exit the screen and make the address change take affect.

8.  It is prudent to put some sort of external label on the device that denotes its address, 

though this can be checked at any time by visiting the “Comm. RS-485” screen.

On units manufactured prior to November 2005 and all units without the standard 

display

, it is necessary to first connect each unit individually (one at a time) to a computer 

and assign a unique address. Then connect all of the units to the Multi-Drop Box. 
Assign each flow or pressure device a unique address as follows:

1. 

Make  sure  the  Multi-Drop  Box  is  connected  to  appropriate  power  and  to  the 

computer.

2.  Connect a DC-62 (or IC-102) cable between the device and any port on the Multi-

Drop Box.

3.  If the unit has a display, check to make sure the device powers up properly.
4.  Make sure ONLY the unit you want to change is connected.
5.  Establish serial communication as outlined in the operating manual for the flow or 

pressure device.

6.  With the flow or pressure device in polling mode, type *@=NEWADDRESS followed 

by the Enter key where NEWADDRESS is any capital letter A thru Z. (e.g. *@=B 

followed by the Enter key)