Alicat BB9 Operating Manual Download Page 5

5

Power

All of the devices attached to the Multi-Drop Box are normally powered through a common 

power bus applied to the terminal block on the front of the box.

When power is supplied through the 4-pin terminal block, the input current must 

not exceed nine amps.

Under some circumstances power may be supplied from a connected device. 

If power 

is being supplied from a connected device, the input current must not exceed  

two amps.

The terminal block requires DC VOLTAGE, appropriate for the devices hooked up to the 

box and of sufficient current to run all of the devices. (See table below.) 
Positive DC voltage must be applied to the socket labeled “POWER” and the power 

ground  must  be  applied  to  the  socket  labeled  “GROUND”.  Connect  the  wires  to  the 

terminal block by following the six steps previously listed.

Device

Voltage 

(Vdc)

Current 

(milliamps)

Meters and Gauges (all ranges)

7-30

50 mA each

0-20 SLPM Controllers and PC, PCD Series Pressure 

Controllers 

12-30

150 mA each

20+ to 100 SLPM Controllers

24-30

750 mA each

100+ to 1500 SLPM Controllers and PCR Series 

Pressure Controllers

24-30

750 mA each

CAUTION! Do not apply AC voltage to the Multi-Drop Box terminal block or to 

individual units as immediate damage will occur!

Do Not

 power units individually through their AC/DC adapter jacks. Potential differences 

between units could damage the power bus in the Multi-Drop Box or damage individual 

units which may be supplied too much voltage. 