Final Assembly QC CON’T
12)
Press the
STORE
button.
Verify
that the
Channel A
signal AND
Channel B
signal LEDs turn on,
and that when the button is held, that the LED pattern is in working order
as in the Front Panel Sub Assembly Test.
Problems
with no response or flickering LEDs indicates poor or failed
mechanical fit for the switch, and the test has FAILED. Problems with the
appearance of the LEDs also indicates the test has FAILED.
Failure
of this inspection indicate that the DUT cannot continue further
testing.
13)
Press the
POWER
button.
Verify
that the unit turns OFF.
Problems
with this inspection indicate that the mechanical fit of the
Power button has FAILED.
Failure
of this inspection indicate that the DUT cannot continue further
testing.
14)
Press the
POWER
button.
Verify
that the unit turns ON.
Problems
with this inspection indicate that the mechanical fit of the
Power button has FAILED.
Failure
of this inspection indicate that the DUT cannot continue further
testing.
15)
Unplug the power supply to the DUT. Then unplug all other cabling to the DUT.
This is the end of the Final Assembly QC test.
Summary of Contents for DEQ230
Page 9: ...WIRING DIAGRAM US EU UK AUS NZ D SERVICE DEQ 230 DOC...
Page 10: ...Packing Diagram US EU AK AUS NZ...
Page 11: ...D SERVICE DEQ 230 DOC Exploded View US AU UK AUS NZ...
Page 28: ...ME2FP208a sch 1 Tue Sep 17 13 07 36 2002...
Page 29: ...ME2FP208a sch 2 Tue Sep 17 13 07 47 2002...
Page 30: ...ME2FP208a sch 3 Tue Sep 17 13 07 56 2002...
Page 31: ...ME2FP208a sch 4 Tue Sep 17 13 08 05 2002...
Page 32: ...ME2FP208a sch 5 Tue Sep 17 13 08 20 2002...
Page 33: ...ME2FP208a sch 6 Tue Sep 17 13 08 28 2002...
Page 34: ......
Page 35: ......
Page 36: ......
Page 37: ......
Page 38: ......