Alemite 3579 Service Manual Download Page 2

SER 3579

Electronic Meter

Revision (3-10)

2

Alemite, LLC

Meter Programming

IMPORTANT: The meter must be in Standby to 
begin any programming function. See Figure 1.

Change Unit of Measure

  1. Press and hold the 

Cal

 and 

Reset

 buttons at the same time.

• The 

Unit

 screen appears.

  2. Repeatedly press the 

Reset

 button.

• The screen scrolls through the available units of measure 

(see 

Figure 1

).

Once the desired units of measure appear:

  3. Press and hold the 

Cal

 button.

• The chosen units of measure are stored in memory and 

active.

• The meter cycles to Standby.

Calibration Factor Determination

To determine the meter’s calibration factor* in use:

  1. Press and hold the 

Cal

 button.

• The calibration factor screen appears.

The word 

Fact

 (short for factory) indicates the calibration set 

at the factory is in use (1.000). 

User

 indicates the 

Fact

ory

 

calibration factor has been superseded and is in use.

If the meter indicates 

User

 but the 

Fact

ory

 

calibration factor 

(1.000) is required:
  2. Press the 

Reset

 button.

• The screen changes from the 

User

 calibration factor to the 

Fact

ory

 

calibration factor.

  3. Press the 

Cal

 button.

• The meter cycles to Standby

.

• The 

Fact

ory calibration factor (1.000) is active.

• The 

User

 calibration factor is deleted from memory.

* The calibration factor is a value that the meter uses to 

calculate the amount of fluid dispensed.

User Calibration

Variables that cause a meter to require recalibration are:

• fluid viscosity
• fluid flow rate
• back-pressure

Field Calibration

This method of programming the meter requires dispensing a 

known quantity of fluid.

  1. Press and hold the 

Cal

 button.

• The calibration factor screen appears.

  2. Press and hold the 

Reset

 button.

• The 

Field

 screen appears.

  3. Dispense any amount of fluid greater than 5 quarts or liters 

into an appropriate-sized graduated beaker. 
• Dispense the fluid at the normal flow rate.

Should the field value match the dispensed value:

  4. Press the 

Reset

 button.

• An upward arrow appears.

  5. Press and hold the 

Reset

 button.

• The meter cycles to Standby.

The calibration factor remains the same.

If the field value does not match the actual amount:

  6. Press the 

Reset

 button.

• An upward arrow appears.

This arrow enables the user to increase the field value.

  7. Press the 

Reset

 button again to change the direction of the 

arrow.
• A downward arrow appears.

This arrow enables the user to decrease the value.

With the proper arrow chosen:

  8. Press the 

Cal

 button.

• The value changes one digit.

NOTE

: Press and hold the 

Cal

 button to change 

the value at a faster rate.

Once the required value displays:

  9. Press and hold the 

Reset

 button.

• The 

User

 calibration factor is stored in memory and active.

• The meter cycles to Standby.

Direct Calibration

This method of calibration is useful to correct an inaccuracy 

of a known percentage. 

For example, if the calibration factor is set at 1.000 and the 

meter registers more than the actual amount dispensed by 5%:

• the calibration factor should be decreased to 0.950

(1.000 x 0.95 = 0.950). 

Conversely, if the meter registers less that the actual amount 

dispensed by 5%:

• the calibration factor should be increased to 1.050

(1.000 x 1.05 = 1.050). 

To program a calculated calibration factor into the meter:

  1. Press and hold the 

Cal

 button.

• The calibration factor screen appears.

  2. Press and hold the 

Reset

 button.

• The 

Field

 screen appears.

  3. Press and hold the 

Reset

 button once again.

• The 

Direct

 screen appears with an upward arrow.

This arrow enables the user to increase the direct value.

  4. Repeat steps 

7

 through 

9

 above.

Reviews: