Alcatel Vacuum Technology France - ASM 182 TD+ - ASM 192 T2D+ User’s Manual
Basic internal calibration of the leak
detector
Edition 03 - December 01
GB 00969 -
Internal
auto-calibration
procedure
Note:
Internal auto-calibration is ON (ON is the setting by
default.)
Different screens appear
during the auto-calibration
giving internal parameters
values.
Audio messages inform the
operator about internal auto-
calibration process during this
one.
When auto-calibration is
complete, the unit is ready to
start a cycle.
The digital voice gives to the
operator the message
“
Detector ready for cycle
”.
Note:
It is possible to start a test cycle after an autocali-
bration failure.
F1
F3
F2
F4
READY FOR CYCLE
INLET ...
STDBY
...
F1
F3
F2
F4
CALIBRATION VALUES:
TEMPERATURE: 29°C
TARGET VAL: 1.27E-07
PLEASE WAIT
...
...
On request internal
auto-calibration
procedure
Note:
Internal auto-calibration is ON (ON is the setting by
default).
Then, the procedure is the same as for the internal auto-
calibration.
F1
F3
F2
F4
READY FOR CYCLE
INLET ...
STDBY
AUTO CAL