44
ATS850 User Manual
Number of channels
2 simultaneously sampled
Maximum Sample Rate
50 MS/s single shot
Minimum Sample Rate
10 KS/s single shot
Full Scale Input ranges
±
20mV,
±
40mV,
±
50mV,
±
80mV
, ±
100mV,
±
200mV,
±
400mV,
±
500mV,
±
800mV,
±
1V,
±
2V,
±
4V,
±
5V,
±
8V,
±
10V and
±
20V, software
selectable.
DC accuracy
±
2% of full scale in all input ranges
Input coupling
AC or DC, software selectable
Input impedance
50
Ω
or 1M
Ω
±
1% in parallel with 30 pF
±
10pF,
software selectable
Input protection
1M
Ω
±
28V (DC + peak AC for CH A,
CH B and EXT only without external
attenuation)
50
Ω
±
8V (DC + peak AC for CH A,
CH B and EXT only without external
attenuation)
Acquisition Memory System
Acquisition Memory/channel Up to 262,140 samples per channel
Record Length
Software selectable with 4-point resolution.
Record length must be a minimum of 256
points. Maximum record length is limited by
the acquisition memory per channel.
Number of Records
Software selectable from a minimum of 1 to a
maximum of 1,000 or (Acquisition Memory Per
Channel / (Record 4)), whichever is
lower
Pre-trigger depth
0 to (Record Length-64), software selectable
with 4 point resolution
Post-trigger depth
Record Length - Pre-trigger depth
Timebase System
Timebase options
Internal Clock or
External Clock (Optional)
Internal Sample Rates
50 MS/s, 25 MS/s, 10 MS/s, 5 MS/s,
2 MS/s, 1 MS/s, 500 KS/s,
200 KS/s, 100KS/s, 50 KS/s,
20KS/s, 10KS/s
Internal Clock accuracy
±
100 ppm
Summary of Contents for ATS850
Page 2: ......
Page 18: ...10 ATS850 User Manual SyncBoard 4X for up to 4 digitizers SyncBoard 8X for up to 8 digitizers...
Page 55: ...ATS850 User Manual 47...
Page 56: ...48 ATS850 User Manual...
Page 57: ......