Akytec PD111 User Manual Download Page 5

 

Specification 

 

 

akYtec GmbH · Vahrenwalder Str. 269 A · 30179 Hannover · Germany · Tel.: +49 (0) 511 16 59 672-0 · www.akytec.de  

 

 

4

 

3.  Specification 

Table 3.1   Specification 

Measuring ranges 

 

 

Nominal pressure, bar 

0...-1 

0...-0,6 

0...-0,4 

0...-0,25 

0...-0,16 

 

Max. working pressure (MWP), 
bar 

2...-1 

2...-1 

2...-1 

0,7...-0,7 

0,7...-0,7 

 

Burst pressure (OPL*), bar

 

1,4 

1,4 

 

Nominal pressure, bar 

0...0,16 

0...0,25 

0...0,4 

0…0,6 

0...1 

 

Max. working pressure (MWP), 
bar 

0,7 

0,7 

 

Burst pressure (OPL*), bar

 

1,4 

1,4 

 

Nominal pressure, bar 

0...1,6 

0...2,5 

0...4 

0...6 

0...10 

 

Max. working pressure (MWP), 
bar 

10 

10 

10 

20 

20 

 

Burst pressure (OPL*), bar

 

20 

20 

20 

40 

40 

 

Nominal pressure, bar 

0...16 

0…25 

0...40 

0...60 

0...100 

 

Max. working pressure (MWP), 
bar 

60 

60 

200 

200 

200 

 

Burst pressure (OPL*), bar

 

120 

120 

400 

400 

400 

Vacuum resistance

 

 

 

P

N

 

≥ 1 bar

 

unlimited vacuum resistance 

P

N

 < 1 bar 

on request 

Power supply 

12...36 V DC 

Current consumption, max. 

70 mA 

Output signal 

4-20 mA 

Accuracy

 

 

 

standard 

± 0.5 % FSO 

option** 

± 0.25 % FSO 

Permissible load 

0...1200 ohm 

Influence effects

 

 

 

supply 

0.01 % FSO / 10 V 

load 

0.05 % FSO / 1 kohm 

Temperature influence 

≤ 0,1 % / 10

 

°C 

Response time 

≤ 100 ms

 

Short circuit protection 

yes 

Reverse polarity protection 

yes 

Permissible temperatures 

 

 

process media 

-40...+100 °C 

ambient 

-40...+80 °C 

storage 

-40...+80 °C 

IP Code 

IP65 

Electromagnetic compatibility 
(EMC) 

EMI and interference immunity according to EN 

61326 

Shock resistance 

480 g / 1 ms (according to EN 60068-2-27) 

Vibration resistance 

9 g at 25..2000 Hz (according to EN 60068-2-6) 

Materials 

 

 

Process connection 

Stainless steel 1. 4307 

Enclosure 

Stainless steel 1. 4307 

Sealing 

no sealing (laser-welding diaphragm) 

Diaphragm 

Stainless steel 1.4435 

Media wetted parts 

Process connection and diaphragm 

Lifetime 

> 500 000 Horas 

Process connection 

M20x1.5; G1/2”; G1/4" (look Chapter 9) 

Summary of Contents for PD111

Page 1: ...r 269 A 30179 Hannover Germany Tel 49 0 511 16 59 672 0 www akytec de PD111 Pressure transmitter for general industrial application User guide PD111_2018 04_0281_EN All rights reserved Subject to tech...

Page 2: ...over Germany Tel 49 0 511 16 59 672 0 www akytec de 1 Contents 1 Intended use 2 2 Safety 3 3 Specification 4 4 Wiring 6 5 Connector mounting 7 6 Maintenance and cleaning 8 7 Transport and storage 9 8...

Page 3: ...ent in liquid and gaseous media compatible with stainless steel AISI 316L 1 4435 AISI 304L 1 4307 The pressure is converted into an electrical signal 4 20mA Restrictions on process medium process medi...

Page 4: ...voided will result in death or serious injury WARNING WARNING indicates a potentially hazardous situation which if not avoided could result in death or serious injury CAUTION CAUTION indicates a poten...

Page 5: ...d vacuum resistance PN 1 bar on request Power supply 12 36 V DC Current consumption max 70 mA Output signal 4 20 mA Accuracy standard 0 5 FSO option 0 25 FSO Permissible load 0 1200 ohm Influence effe...

Page 6: ...akytec de 5 Electrical connection EN 175301 803 A ISO 4400 Supply 1 Supply 2 Shield Ground contact Mounting position any Dimensions 92 x 52 x 34 mm Weight approx 400 g CE conformity EMC Directive 200...

Page 7: ...separately or screened from the supply cables NOTICE Only a shielded cable may be used for the signal lines Fig 4 1 Wiring diagram Use the diagram Fig 4 2 to select the permissible load resistance Rec...

Page 8: ...using the screw hole or the cable fitting 4 Unscrew the cable fitting 4 and take out the cable seal 3 Lead the cable through the cable fitting 4 and the cable seal 3 in the enclosure 2 Connect the cab...

Page 9: ...ance free If needed the housing of the device can be cleaned using a damp cloth and non aggressive cleaning solutions in switched off state When cleaning the diaphragm the following is to observe The...

Page 10: ...after delivery it must be carefully stored in a safe and protected place Storage temperature range 40 80 C WARNING Remove all the adherent media rests on the process connection before storing the devi...

Page 11: ...ircuit Check the cable line Voltage reversal Reverse polarity Output signal is unsta ble or remains con stant even if pressure is changing Process connection is blocked up Clean the connection Loss of...

Page 12: ...Dimensions akYtec GmbH Vahrenwalder Str 269 A 30179 Hannover Germany Tel 49 0 511 16 59 672 0 www akytec de 11 9 Dimensions...

Page 13: ...4 0 0 M 0 0 6 G 6 0 0 M 0 1 G 1 P 0 B 0 1 6 G 1 P 6 B 0 2 5 G 2 P 5 B 0 4 G 4 P 0 B 0 6 G 6 P 0 B 0 10 G 0 1 0 B 0 16 G 0 1 6 B 0 25 G 0 2 5 B 0 40 G 0 4 0 B 0 60 G 0 6 0 B 0 100 G 1 0 0 B 0 125 0 12...

Reviews: