Akytec PD111 User Manual Download Page 3

 

Intended use 

 

 

akYtec GmbH · Vahrenwalder Str. 269 A · 30179 Hannover · Germany · Tel.: +49 (0) 511 16 59 672-0 · www.akytec.de  

 

 

2

 

1.  Intended use 

The PD111 is a pressure transmitter with internal diaphragm made of stainless steel AISI 
316L and with a silicon measuring cell. The transmitter is applied for gauge pressure 
measurement in liquid and gaseous media compatible with stainless steel AISI 316L / 
1.4435 (AISI 304L / 1.4307). The pressure is converted into an electrical signal 4-20mA.

 

Restrictions on process medium:  
– 

process medium may not corrode the wetted parts of the transmitter 

– 

process medium may not crystallize or harden in process connection 

– 

process medium temperature must be between -40 and +100 °C 

The pressure transmitters described in this guide may be operated only in explosion proof 
areas without aggressive vapors and gases.

 

 

 

Summary of Contents for PD111

Page 1: ...r 269 A 30179 Hannover Germany Tel 49 0 511 16 59 672 0 www akytec de PD111 Pressure transmitter for general industrial application User guide PD111_2018 04_0281_EN All rights reserved Subject to tech...

Page 2: ...over Germany Tel 49 0 511 16 59 672 0 www akytec de 1 Contents 1 Intended use 2 2 Safety 3 3 Specification 4 4 Wiring 6 5 Connector mounting 7 6 Maintenance and cleaning 8 7 Transport and storage 9 8...

Page 3: ...ent in liquid and gaseous media compatible with stainless steel AISI 316L 1 4435 AISI 304L 1 4307 The pressure is converted into an electrical signal 4 20mA Restrictions on process medium process medi...

Page 4: ...voided will result in death or serious injury WARNING WARNING indicates a potentially hazardous situation which if not avoided could result in death or serious injury CAUTION CAUTION indicates a poten...

Page 5: ...d vacuum resistance PN 1 bar on request Power supply 12 36 V DC Current consumption max 70 mA Output signal 4 20 mA Accuracy standard 0 5 FSO option 0 25 FSO Permissible load 0 1200 ohm Influence effe...

Page 6: ...akytec de 5 Electrical connection EN 175301 803 A ISO 4400 Supply 1 Supply 2 Shield Ground contact Mounting position any Dimensions 92 x 52 x 34 mm Weight approx 400 g CE conformity EMC Directive 200...

Page 7: ...separately or screened from the supply cables NOTICE Only a shielded cable may be used for the signal lines Fig 4 1 Wiring diagram Use the diagram Fig 4 2 to select the permissible load resistance Rec...

Page 8: ...using the screw hole or the cable fitting 4 Unscrew the cable fitting 4 and take out the cable seal 3 Lead the cable through the cable fitting 4 and the cable seal 3 in the enclosure 2 Connect the cab...

Page 9: ...ance free If needed the housing of the device can be cleaned using a damp cloth and non aggressive cleaning solutions in switched off state When cleaning the diaphragm the following is to observe The...

Page 10: ...after delivery it must be carefully stored in a safe and protected place Storage temperature range 40 80 C WARNING Remove all the adherent media rests on the process connection before storing the devi...

Page 11: ...ircuit Check the cable line Voltage reversal Reverse polarity Output signal is unsta ble or remains con stant even if pressure is changing Process connection is blocked up Clean the connection Loss of...

Page 12: ...Dimensions akYtec GmbH Vahrenwalder Str 269 A 30179 Hannover Germany Tel 49 0 511 16 59 672 0 www akytec de 11 9 Dimensions...

Page 13: ...4 0 0 M 0 0 6 G 6 0 0 M 0 1 G 1 P 0 B 0 1 6 G 1 P 6 B 0 2 5 G 2 P 5 B 0 4 G 4 P 0 B 0 6 G 6 P 0 B 0 10 G 0 1 0 B 0 16 G 0 1 6 B 0 25 G 0 2 5 B 0 40 G 0 4 0 B 0 60 G 0 6 0 B 0 100 G 1 0 0 B 0 125 0 12...