User Manual of Smart Digital Storage Oscilloscope
48
2.
Press the
H3
menu selection button to set XY Mode
ON
. The display format
is changed to be XY mode (see
Fig.5-33
).
Fig.5-33
XY Display Mode
Cymometer
It is a 6-digit cymometer. The cymometer can measure frequencies from 2Hz to the
full bandwidth. Only if the measured channel has triggering signal and in
Edge
mode,
it can measure frequency correctly. In the
Single
trigger mode , it is a one channel
cymometer and it can only measure the frequency of the triggering channel. In the
ALT
trigger mode, it is a two channel cymometer and it can measure the frequency of
two channels. The cymometer is displayed at the right bottom of the screen.
To turn the cymometer on or off:
1.
Press the
Display
button.
2.
In the
Display
menu, press the
H4
button to toggle between the cymometer
display
ON
or
OFF
.
VGA Output
The VGA port could be connected to a computer monitor. The image of the
oscilloscope can be clearly displayed on the monitor.
To set the VGA Output:
1.
Press the
Display
button.
2.
In the
Display
menu, press the
H5
button to toggle between
ON
or
OFF
.