User Manual of Smart Digital Storage Oscilloscope
60
Measure the frequency, the peak-to-peak voltage of the Channel CH1 and the mean,
the RMS of the Channel CH2, following below steps:
1.
Press the
Measure
button to show the automatic measurement function menu.
2.
Press the
H1
button to display the
Add
menu.
3.
Press the
F2
button and choose
CH1
as the source.
4.
Press the
F1
button, the type items will display at the left of screen, and turn the
M
knob to choose
Period
.
5.
Press the
F4
button, the period options added completes.
6.
Press the
F1
button again, the type items will display at the left of screen, and turn
the
M
to choose
Freq
.
7.
Press the
F4
button, the frequency added completes, finish setting of CH1.
8.
Press the
F2
button and choose
CH2
as the source.
9.
Press the
F1
button, the type items will display at the left of screen, and turn the
M
to choose
Mean
.
10.
Press the
F4
button, the Mean added completes.
11.
Press the
F1
button, the type items will display at the left of screen, and turn the
M
to choose
PK-PK
.
12.
Press the
F4
button, the PK-PK added completes, finish setting of CH2.
The measured value will be displayed in the reading window automatically (see
Fig.5-47).
Fig.5-47 automatic measurement
The automatic measurement of voltage parameters