Table 206
Recommended separation distance between portable and mobile RF communication
equipment and the dental unit
The equipment is intended for use in an electromagnetic environment in which radiated RF
disturbance are controlled. The customer or the user of the equipment can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the equipment as recommended below, according to
the maximum output power of the communications equipment.
Rated
maximum
output power of
transmitter (W)
Separation distance according to frequency of transimitter
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2.5 GHz
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guideline may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
48
P/N: 8013093 Rev
:
F
AJ16 User's Guide