Service the oil collector on the delivery system once a week for
normal usage and more often for heavier use. To service:
1 Release the oil collector cover under the Dentist Unit by turning
counterclockwise and discard the old gauze.
2. Fold a new gauze into quarters and place inside the cover.
3. Tighten the oil collector cover closed by turning clockwise
38
8.3 Clean Oil Collector
8.4 Clean the Filter
Figure 38
The solids collector aids in preventing solids from entering the draining
system.
To ensure proper suction from the vacuum and maintain proper
treatment room asepsis, discard
and replace the solids collector at least twice a week (see Figure 39).
Clean the collector
:
1.
Close the suction system
2.
Swing the fixing frame and open the collector, take out the filter
3.
Dump the waste.
4.
Replace the filter and collector.
Figure 39
P/N: 8013093 Rev
:
F
AJ16 User's Guide