11
The 24 Watt Metal Halide Light
Source Caution!
Never look directly into the light emitting part of the
Videoscope or of the Light Source. There is ultra-
violet and infrared as well as highly intensive visible
light coming from the Light Source. These can
cause injuries of eyes and skin.
Use the Aperture to adjust the intensity of the light.
For most applications we recommend to open it
completely.
For cleaning of the distal end or changing the
Objective switch of the Light Source or remove the
Light Guide Cable from it.
The Metal Halide Lamp installed needs 30 seconds
from engaging to generate full and steady luminance.
An untimely (<30sec.) power-down extremely
reduces the durability of the lamp or can even lead
to its destruction.
High Voltage –
Danger to Life!
Always have a lamp installed into the Light Source
(or even let a depleted lamp installed if necessary)
and close the casing before starting the iTool!
System Presentation
Switch ON/OFF
Aperture