- 6 -
5.1.4 Fill in the distilled water
Open the top lid, and fill the tank with distilled water by cup or tank. When you hear
a beep signal, it means the water level is exceed
the max. Level. Please stop filling immediately.
5.2 Prepare the material to be sterilized
To get a better effectiveness of the sterilization process and to preserve
the material in time, follow the indications below reported.
* Arrange the tools of different metal (stainless steel, moderate steel,
aluminum, etc.) on different trays or however well separate between them;
* In case of not stainless steel tools, interpose a sterilization paper napkin or
muslin cloth between tray and tool, avoiding direct contacts between the
two different materials;
* Verify all the tools are sterilized in open position;
* Arrange the containers (glasses, cups, test-tubes, etc.) on one side or
inverted position, avoiding possible water stagnation;
* Don't overload the trays over the stated limit (see Appendix 1).
* Don't stack the trays one above the other or put them in direct contact with
the walls of the sterilization chamber.
* Always use the instrument tray handle.
* Wrap the tools one by one or, if more tools have to be set in the same wrap,
verify that they are of the same metal;
* Seal the wrap with sterilization adhesive ribbon or by a thermal sealer.