background image

15

FILTER REPLACEMENT

A. To remove the spent ACFHEPA filter:

1.  Turn the unit on and make sure that the blower is running. The safety filter, located in the top of the 

workstation, will protect you from any residual powders that might come off of the saturated primary 

HEPA filter while replacing it.

2.  Place 2 appropriate poly bags on the base of the workstation. You will place the saturated prefilter and 

HEPA filter inside the bag and seal it before removing it from the hood.

3.  Don the proper PPE for handling the substance that have been used inside the hood (refer to the 

MSDS sheets). Remove the 4 black knobs holding the prefilter grid in place. This will expose the 

saturated pre-filter. 

DO NOT REMOVE THE PRE-FILTER FROM THE HOOD!

4.  Remove the pre-filter from the pre-filter grid and place it inside the poly bag. Wipe any residue off of 

the pre-filter grid and base of the workstation. Place the rag inside the bag with the pre-filter and seal 

the bag.  

5.  With a small wrench, remove the ¼ - 20 hex nuts and remove the filter hold down.

6.  Place the spent HEPA filter into the second poly bag.  Wipe any residue off of the filter hold down 

frame and any other obvious areas. Place the second rag into the bag and seal the bag. Dispose of 

both bags in accordance with your local guidelines for the substances that have been used inside the 

hood.

B.  To install the new ACFHEPA filter:

1.  Remove the ACFHEPA from the packaging. Visually inspect the HEPA filter for rips in the filter media.

2.  Place the filter into the filter grid with the gasket and screen side facing away from you.

3.  Place the filter hold down frame over the HEPA filter and secure it with the hex nuts.

4.  Place the new pre-filter into the pre-filter grid and secure into place with the 4 black knobs.

5.  Dispose of your PPE in accordance with your local guidelines for the substances that have been used 

inside the hood. 

 

 

Check the display screen to assure that the enclosure is functioning properly:

1. If the Low Airflow Alarm is not activated, proceed to use the enclosure.

2. If the Low Airflow Alarm is activated, contact our Technical Support Department at (800)849-0472. 

Direct dial (919)255-3220. Monday - Friday 8:00AM - 5:00PM EST.

MAINTENANCE

Summary of Contents for POWDERSAFE AC794BE

Page 1: ...Fax 919 255 6120 Website www aircleansystems com Email info aircleansystems com Product designs are subject to change without notice 2013 AirClean Systems Inc ACOM1207PS BE rev1 OPERATOR S MANUAL POW...

Page 2: ...rating Instructions for Fumes and Vapors and Alarm Messages 7 Airflow Path 8 Side View with Disposal Port 9 View of Base with Bag Access and Collar 10 Controller Instructions 11 Programming Menus 11 1...

Page 3: ...was required to protect applications from carbon dust AirClean Systems has eliminated this problem with breakthrough bonded carbon technology Bonded carbon filters are solid blocks of activated cocon...

Page 4: ...FHEPA ACFHEPA ACFHEPA Dimensions 12 x 24 x 3 12 x 24 x 3 12 x 24 x 3 Requirements 3 4 4 Secondary HEPA Filter3 Type ACFHEPA ACFHEPA ACFHEPA Dimensions 12 x 24 x 3 12 x 24 x 3 12 x 24 x 3 Requirements...

Page 5: ...on the cart or the counter C Remove the prefilter for inspection before use Refer to the MAINTENANCE section of this manual for complete instructions D Connect the supplied power cord to the unit and...

Page 6: ...ide your enclosure B When working within the enclosure keep the front sash down and perform your procedure in the center to rear of the enclosure C If the low airflow alarm sounds and or lights up the...

Page 7: ...es please call our office to verify compatibility before use in this case your filter may have another model number Direct Dial 919 255 3220 or 800 849 0472 Monday Friday 8 00 5 00 PM EST 2 Connect th...

Page 8: ...re filtration and HEPA filtration 3 Particulate free air moves upward to C where it enters the HEPASafe filtration area This filtration is only challenged with particulate during primary filter change...

Page 9: ...9 Disposal Port Rubber Band...

Page 10: ...ace desired barrel on lift and center in hole of workstation 3 Raise lift so barrel breaks plane of base at a comfortable height 4 Pull top of bag out of barrel and drape around polycarb ring 5 Secure...

Page 11: ...POWER Turns on off power to the controller and its keys This DOES NOT cut off power to the entire enclosure The main power switch is located above the power inlet on the back right hand side of the en...

Page 12: ...larm when it sounds by touching the alarm button on the controller The visible alarm will still work FILTER This menu shows the serial number and the purchase date of the filter For every new filter i...

Page 13: ...usly display face velocity while blower is activated PUMP Inactive key for enclosure LIGHT One touch to enable fluorescent light if installed TIMER User definable lab event timer AUX Inactive key for...

Page 14: ...and pre filter grid are located at the back of the unit 2 Look through the slats on the pre filter grid If the prefilter is discolored the original color is green replacement is needed B To replace th...

Page 15: ...uts and remove the filter hold down 6 Place the spent HEPA filter into the second poly bag Wipe any residue off of the filter hold down frame and any other obvious areas Place the second rag into the...

Page 16: ...Repeat steps above for holes 2 and 3 4 For standard setup complete assembly using hole 4 For seated use complete assembly using hole 5 5 Tighten all bolts completely 6 Installation of casters should b...

Page 17: ...Cross Piece 2 Locking Caster 2 Standard Caster 4 Washer 8 1 Washer 4 1 x Bolt 8 2 x Bolt Standard Setup For Seated Use Mounting Holes Hole 3 Hole 5 Hole 4 Hole 1 Hole 2 ASSEMBLED CART CONFIGURATIONS...

Page 18: ...uorescent Fixture 30w 1 10 50428 Fuse 6 3a 250V Slo Blo 1 1 1 06 50481 Gas Strut 60 2 2 06 51087 Gas Strut 90 2 06 50012 Hinge Door Positioning 6 8 12 06 50053 Hinge Medium Black 3 4 5 11 52756 Keypad...

Page 19: ...ou the product If the cause is determined to be a manufacturing fault the dealer or AirClean Systems will repair or replace all defective parts to restore the unit to operation Under no circumstances...

Page 20: ...ery NOTE Do not return any materials without authorization from AirClean Systems Products returned without prior authorization will not be accepted AirClean Systems and its dealers are not responsible...

Page 21: ...Woodard Drive Raleigh NC 27604 Phone 919 255 3220 Toll Free 800 849 0472 Fax 919 255 6120 Visit AirClean Systems on the Internet at http www aircleansystems com Email AirClean Systems at info airclean...

Page 22: ...aleigh NC 27604 Phone 919 255 3220 Toll Free 800 849 0472 Fax 919 255 6120 E mail info aircleansystems com Product designs are subject to change without notice 2013 AirClean Systems Inc ACOM1207PS BE...

Reviews: