6
SHIPMENT CONTENTS
• AirClean
®
Systems PowderSafe™ enclosure complete
®
®
• Operator’s Manual
• Power Cord
• Optional Equipment (if ordered)
INSTALLATION AND ASSEMBLY INSTRUCTIONS
Caution: This enclosure should not be installed near doors, windows or air conditioning/
heating/ ventilation systems. Air must be allowed to circulate freely around the enclosure.
Allow at least 3” between the top of the enclosure and any overhead cabinets or shelving for
adequate exhausting.
A. If ordered, assemble the mobile cart. Installation instructions are included with the cart and in
the “OPTIONS” section of this manual.
B. Place the enclosure on the cart or the counter.
C. Remove the prefi lter for inspection before use. Refer to the “MAINTENANCE” section of this
manual for complete instructions.
D. Connect the supplied power cord to the unit and to the power outlet. Flip the main power
switch of the unit above the power inlet module.
(Power cord connection located on the top of the enclosure)
This will activate the main power. The screen on the controller should light up and display the
time and the date.
E. Press the “Power” button on the controller. This will activate the controls for the enclosure. If
the blower does not activate press the quick key labeled, “Blower”. If the blower still does not
activate contact the technical support staff at the factory. Direct Dial 919-255-3220 or 800-849-
0472. Monday-Friday 8:00-5:00 PM EST.
INSTALLATION AND ASSEMBLY