5
TECHNICAL SPECIFICATIONS
SPECIFICATIONS
AC765C
AC775C
AC785C
Physical Dimensions
Outside Dimensions
36”W x 30”D x 54”H 48”W x 30”D x 54”H 72”W x 30”D x 54”H
Inside Dimensions
35”W x 25”D x 35”H 47”W x 25”D x 35”H 71”W x 25”D x 35”H
Weight, including fi lters
Airfl ow Characteristics
Low Airfl ow Alarm
AirSafe™
AirSafe™
AirSafe™
Volume of Filtered Air
660 CFM
660 CFM
660 CFM
Linear Face Velocity
Variable
Variable
Variable
Electrical
Voltage
110 V or 220 V
110 V or 220 V
110 V or 220 V
Current
5.6 A / 2.8 A
5.6 A / 2.8 A
5.6 A / 2.8 A
Power Consumption
616 W
616 W
616 W
Operation
Speed Selection
Auto / Manual
Auto / Manual
Auto / Manual
Construction
Front Panel
Powder Coated Steel Powder Coated Steel Powder Coated Steel
Base and Sides
Polypro
Polypro
Polypro
Sash Material
Polycarbonate
Polycarbonate
Polycarbonate
Pre-fi lter
1
Type
ACFPRE1
ACFPRE1
ACFPRE1
Dimensions
12” x 24” x 1”
12” x 24” x 1”
12” x 24” x 1”
Requirements
2
3
4
Primary HEPA Filter
2
Type
ACFHEPA
ACFHEPA
ACFHEPA
Dimensions
12” x 24” x 3”
12” x 24” x 3”
12” x 24” x 3”
Requirements
2
3
4
Secondary HEPA Filter
3
Type
ACFHEPA
ACFHEPA
ACFHEPA
Dimensions
12” x 24” x 3”
12” x 24” x 3”
12” x 24” x 3”
Requirements
2
3
4
Bonded Carbon Filter
4
ilter
ilter (Optional)
Type
(Application Based)
(Application Based)
(Application Based)
Dimensions
12” x 24” x 3”
12” x 24” x 3”
12” x 24” x 3”
Requirements
2
3
4
NOTE 1 -
Your enclosure is shipped with the prefi lter(s) attached to the inside of the enclosure. AirClean
®
Systems
®
®
prefi lters are fi ltrete
electrostatic particulate fi lters with 65% effi ciency for particulate down to 0.5 microns in diameter (DOP test). The pre-fi lter
must be checked on a scheduled basis. Refer to the “MAINTENANCE” section of this manual.
NOTE 2 -
Your AirClean
®
Systems enclosure was shipped with the Primary and Secondary HEPA fi lters installed.
®
®
NOTE 3 -
If your application is particulate-based only, a HEPA fi lter is required. The HEPA fi lters
will not protect the
will not protect the operator f
operator from
any fumes
y fumes
.
NOTE 4 -
Bonded carbon fi lters are required for chemical-based applications or when trying to contain aromas from powders. The
bonded carbon fi lter will not aid in the containment of particulate.