   CLEANING
 
 

Exterior:

 

• Clean with only a soft, damp cotton cloth or an acrylic cleaner. 

 

• 

Do not use paper products to clean unit, as they may scratch the material.

 

Interior:

 

• A high level disinfectant or 10% bleach/water solution may be used as needed. Sporicidin

®

 is

 

  an approved disinfectant for use with AirClean

®

 Systems PCR workstations.

WARNING

Interior  cleaning  with  an  approved  disinfectant  should  be 

completed  routinely.  Residual  moisture  must  be  removed  or 

UV radiation may cause chemical burn of the polycarbonate. If 

this occurs contact AirClean

®

 Systems technical support before 

continued usage of product.

CATALOG NO.

DESCRIPTION

AC632 AC648

77-52398

Electronics Box (648)

N/A

1

11-52757

Controller Touch Pad

1

1

10-50055

Power Inlet Fuse, 3.15 A, Slow Blow

1

N/A

10-50428

Power Inlet Fuse, 6.3 A, Slow Blow

N/A

1

03-50184

IEC Connector

1

1

77-52411

Door Switch

1

1

77-52441

Light Fixture

2

4

01-52053

Power Cord

1

1

09-50031

¼-20 x 1 PH bolt for fixed sash

4

4

09-50547

¼-20 Lock nut

4

4

06-50318

Sash Hinges

4

8

77-52431

Polycarbonate Front Sash, Complete

1

N/A

77-52407

Polycarbonate Front Sash, Complete

N/A

1

77-53715

Polycarbonate shell

1

N/A

77-52409

Polycarbonate shell

N/A

1

73-63281

Polypropylene Base

1

N/A

73-64887

Polypropylene Base

N/A

1

04-50647

Polycarbonate Sash Rails

2

2

10-50215

15W Cool White Fluorescent Lamp

1

2

ACAUVBL4

UV Bulbs, Pack of 4

1

2

SPARE PARTS LIST

Product designs are subject to change without notice, contact Customer Service before placing order.

Summary of Contents for AC600 Series

Page 1: ...e 800 849 0472 Fax 919 255 6120 Website www aircleansystems com Email info aircleansystems com Product designs are subject to change without notice 2012 AirClean Systems OPERATOR S MANUAL DEAD AIR WOR...

Page 2: ...mming Menus 12 Operating Instructions 13 Alarm Messages 13 Maintenance Bulbs 14 16 Cleaning 16 Spare Parts List 16 Options 17 18 Shipping Claims 19 Contact Information 19 FIGURES PAGE Figure 1 AC600DB...

Page 3: ...dealer or AirClean Systems will repair or replace all defective parts to restore the unit to operation Under no circumstances shall AirClean Systems be liable for indirect consequential or special dam...

Page 4: ...61000 4 2 Electromagnetic Compatibility ESD IEC 61000 4 3 Electromagnetic Compatibility Radiated Immunity IEC 61000 4 4 Electromagnetic Compatibility Fast Transient IEC 61000 4 5 Electromagnetic Compa...

Page 5: ...and certified as complying with the applicable Underwriter s Laboratory USA regulations and recommendations The presence of this symbol means that this equipment has been designed tested and certifie...

Page 6: ...le Always replace fuses with those specified These instructions must be placed in an easily seen location DO NOT use this equipment for purposes other than nucleic acid preparation and amplification o...

Page 7: ...Type 3A 250V Slow Blow 5 X 20 6 3A 250V Slow Blow 5 X 20 Requirements 1 110V 2 230V 1 110V 2 230V Environmental Usage Indoor Use Only Indoor Use Only Altitude Up to 2000m Up to 2000m Humidity 80 Max u...

Page 8: ...fe of 1000 hours in your workstation The UVTect Controller tracks UV bulb hours and alerts the user when to replace the UV bulb The overlapping sash design prevents operator exposure to shortwave UV e...

Page 9: ...e to lift from the bottom of the workstation See figure 2 CAUTION DO NOT lift the workstation by the head assembly the sash or any protruding wires 4 Place the workstation in the desired location Once...

Page 10: ...the workstation See figure 4 10 Locate the power entry module on the back of the workstation Insert the female end of the linecord into the power entry module and the male end into an appropriate elec...

Page 11: ...and turn off when sash is closed TIMER User definable lab event timer AUX Inactive key for workstation NAVIGATION KEYS MODE Allows the operator to move between different program screens SELECT Allows...

Page 12: ...set a specific amount of time Pressing the Timer quick key on the controller will activate the timer UV Allows operator to set operation time of UV light It is factory preset for 15 minutes TIME DATE...

Page 13: ...e UV light timer will count down from 15 minutes The UV light will automatically turn off when the timer reaches zero Note The UV light will automatically turn off when the front sash is opened and wi...

Page 14: ...ompletely open the folding front sash by carefully resting it on the head assembly of the workstation 3 Using both hands turn the bulb counter clockwise turn and remove from the light socket See figur...

Page 15: ...ly open the folding front sash by carefully resting it on the head assembly of the workstation 3 Using both hands turn the bulb counter clockwise turn and remove from the light socket 4 Take note of t...

Page 16: ...648 77 52398 Electronics Box 648 N A 1 11 52757 Controller Touch Pad 1 1 10 50055 Power Inlet Fuse 3 15 A Slow Blow 1 N A 10 50428 Power Inlet Fuse 6 3 A Slow Blow N A 1 03 50184 IEC Connector 1 1 77...

Page 17: ...le 4 For seated use complete assembly using hole 5 5 Tighten all bolts completely 6 Installation of casters should be per illustration on page 21 MOUNTING OF Enclosure TO ASSEMBLED CART 1 Remove the a...

Page 18: ...g Caster 2 Standard Caster Hook Loop Fasteners 4 Hook 4 Loop Standard Setup For Seated Use Location for Hook Loop Fasteners Hole 3 Hole 5 Hole 4 Hole 1 Hole 2 ASSEMBLED CART CONFIGURATIONS CART COMPON...

Page 19: ...rwise the damage claim may be refused If concealed damage or pilferage is discovered notify the carrier immediately and retain the entire shipment intact for inspection Interstate Commerce Commission...

Page 20: ...rive Raleigh NC 27604 Phone 919 255 3220 Toll Free 800 849 0472 Fax 919 255 6120 Website www aircleansystems com Email info aircleansystems com Product designs are subject to change without notice 201...

Reviews: