background image

15

  

To replace the fluorescent light bulb in the AC648DBC model:

 

 

 

 

 

 

 

1. Disconnect linecord from the power entry module. Disconnect the door switch cable and

 

    light cables. See figure 4. (Page 10)

 

2. Using both hands, turn the bulb counter clockwise ¼ turn and remove the bulb

 

    from the socket. 

 

3. Install new bulb by sliding into the light socket and turning it clockwise ¼ turn until the

 

    bulb locks into place.

 

4. Reconnect the linecord, door switch cable, and light cables.

 

UV Bulb Replacement

 

Each UV bulb has a life expectancy of 1000 bulb hours. The UVTect™ Controller will track the

 

remaining UV bulb hours. Each time the UV bulb is illuminated, a minimum of one hour will be

 

used. The alarm will sound when 100 UV bulb hours remain. When the life of the UV bulb has

 

expired, the UV light will not work until a new bulb is installed. 

 

 

 

 

 

To replace the UV light bulb(s) in the AC632DBC or AC648DBC models:

 

1. Turn off power at the power entry module. Disconnect linecord from the power entry 

   

 

   

    

module. See figure 4. (Page 10)

 

2. Completely open the folding front sash by carefully resting it on the head assembly of

 

    the workstation.

 

3. Using both hands, turn the bulb counter clockwise ¼ turn and remove from the light socket.

 

4. Take note of the serial number on the sleeve of the new bulb prior to installing. Install new

 

    bulb by sliding into the light socket and turning clockwise ¼ turn until the bulb locks

 

5. Reconnect linecord to power entry module. Turn the power on at the power entry module. 

          

 

   

See figure 4. (Page 10)

 

To enter validation number of new UV bulb:

 

1. Press the “MODE” arrow keys until the UV Lamp screen is displayed. This screen will

 

    display the number of remaining useful hours for the UV bulb.

 

2. Press the “SELECT” up arrow key. 

 

3. Using the “SET” keys, enter the new UV bulb validation number starting with the first digit. 

 

    Press the “SELECT” up arrow key to move to the next digit.

 

4. Repeat until the entire validation number is entered.

 

5. Press the “SELECT” up arrow key until the “<” caret blinks.

 

6. Finish by pressing the “SET” up arrow key once to store the UV bulb validation number.

CAUTION

Disconnect the linecord from the power entry 

module prior to proceeding.

CAUTION

Disconnect the linecord from the power entry 

module prior to proceeding.

Summary of Contents for AC600 Series

Page 1: ...e 800 849 0472 Fax 919 255 6120 Website www aircleansystems com Email info aircleansystems com Product designs are subject to change without notice 2012 AirClean Systems OPERATOR S MANUAL DEAD AIR WOR...

Page 2: ...mming Menus 12 Operating Instructions 13 Alarm Messages 13 Maintenance Bulbs 14 16 Cleaning 16 Spare Parts List 16 Options 17 18 Shipping Claims 19 Contact Information 19 FIGURES PAGE Figure 1 AC600DB...

Page 3: ...dealer or AirClean Systems will repair or replace all defective parts to restore the unit to operation Under no circumstances shall AirClean Systems be liable for indirect consequential or special dam...

Page 4: ...61000 4 2 Electromagnetic Compatibility ESD IEC 61000 4 3 Electromagnetic Compatibility Radiated Immunity IEC 61000 4 4 Electromagnetic Compatibility Fast Transient IEC 61000 4 5 Electromagnetic Compa...

Page 5: ...and certified as complying with the applicable Underwriter s Laboratory USA regulations and recommendations The presence of this symbol means that this equipment has been designed tested and certifie...

Page 6: ...le Always replace fuses with those specified These instructions must be placed in an easily seen location DO NOT use this equipment for purposes other than nucleic acid preparation and amplification o...

Page 7: ...Type 3A 250V Slow Blow 5 X 20 6 3A 250V Slow Blow 5 X 20 Requirements 1 110V 2 230V 1 110V 2 230V Environmental Usage Indoor Use Only Indoor Use Only Altitude Up to 2000m Up to 2000m Humidity 80 Max u...

Page 8: ...fe of 1000 hours in your workstation The UVTect Controller tracks UV bulb hours and alerts the user when to replace the UV bulb The overlapping sash design prevents operator exposure to shortwave UV e...

Page 9: ...e to lift from the bottom of the workstation See figure 2 CAUTION DO NOT lift the workstation by the head assembly the sash or any protruding wires 4 Place the workstation in the desired location Once...

Page 10: ...the workstation See figure 4 10 Locate the power entry module on the back of the workstation Insert the female end of the linecord into the power entry module and the male end into an appropriate elec...

Page 11: ...and turn off when sash is closed TIMER User definable lab event timer AUX Inactive key for workstation NAVIGATION KEYS MODE Allows the operator to move between different program screens SELECT Allows...

Page 12: ...set a specific amount of time Pressing the Timer quick key on the controller will activate the timer UV Allows operator to set operation time of UV light It is factory preset for 15 minutes TIME DATE...

Page 13: ...e UV light timer will count down from 15 minutes The UV light will automatically turn off when the timer reaches zero Note The UV light will automatically turn off when the front sash is opened and wi...

Page 14: ...ompletely open the folding front sash by carefully resting it on the head assembly of the workstation 3 Using both hands turn the bulb counter clockwise turn and remove from the light socket See figur...

Page 15: ...ly open the folding front sash by carefully resting it on the head assembly of the workstation 3 Using both hands turn the bulb counter clockwise turn and remove from the light socket 4 Take note of t...

Page 16: ...648 77 52398 Electronics Box 648 N A 1 11 52757 Controller Touch Pad 1 1 10 50055 Power Inlet Fuse 3 15 A Slow Blow 1 N A 10 50428 Power Inlet Fuse 6 3 A Slow Blow N A 1 03 50184 IEC Connector 1 1 77...

Page 17: ...le 4 For seated use complete assembly using hole 5 5 Tighten all bolts completely 6 Installation of casters should be per illustration on page 21 MOUNTING OF Enclosure TO ASSEMBLED CART 1 Remove the a...

Page 18: ...g Caster 2 Standard Caster Hook Loop Fasteners 4 Hook 4 Loop Standard Setup For Seated Use Location for Hook Loop Fasteners Hole 3 Hole 5 Hole 4 Hole 1 Hole 2 ASSEMBLED CART CONFIGURATIONS CART COMPON...

Page 19: ...rwise the damage claim may be refused If concealed damage or pilferage is discovered notify the carrier immediately and retain the entire shipment intact for inspection Interstate Commerce Commission...

Page 20: ...rive Raleigh NC 27604 Phone 919 255 3220 Toll Free 800 849 0472 Fax 919 255 6120 Website www aircleansystems com Email info aircleansystems com Product designs are subject to change without notice 201...

Reviews: