FUSES 

Switched units: A fuse is provided on the switch front plate. 

Speed controlled units: An additional fuse is provided within the speed control panel. To check this fuse, 

remove the service panel; the speed control panel is mounted on the rear of the service panel. 

Compressed Air 

VertEx VP requires a compressed air supply at 5 bar. Supplied air must be clean, and oil/moisture free. 
Contaminated air will result in poor cleaning of filters, filter degradation, or failure of components. We 

recommend installation of a shut-off valve, bleed-type regulator with gauge, filter, and condensate filter prior to 
the air inlet to the AirBench. 

You should regularly check the tank to ensure it is clean and dry internally. If moisture is present inside the 
tank, this can lead to corrosion. Your insurer may also wish to access the tank for regular inspections and you 
should ensure they are aware of the presence of an air receiver of 5l volume within this unit. 

Ducted Units 

As standard, VertExes are configured to discharge upwards through the top of the unit, without ductwork. 
Ducting should not be retrofitted to the discharge from the unit without consultation with the manufacturer. 

When configured for exhaust to atmosphere via ductwork a discharge spigot will be visible. These units are 

designed to connect to ductwork of the same size as the discharge spigot (generally 250 or 300 mm 
diameter). 

Ducting should be installed by a professional and should not provide greater than 100 Pascals resistance. 

Flexible ducting should be avoided as it may cause excessive back-pressure on the fan, and lead to reduced 
airflow. 

Any other configuration is non-standard and should be discussed with the manufacturer. 

Assembly of Accessories 

These items apply only when they have been ordered with the VertEx. 

ENCLOSURE 

Enclosures are supplied flat packed for assembly at point of use. Each enclosure is supplied with installation 

drawings and assembly should follow the process laid out in these drawings. 

© Copyright AirBench Ltd 2016

Page 

!

 of 

!

6

16

Summary of Contents for Vertex VP

Page 1: ...rtex VP Operation and Maintenance This Edition 2 October 2018 AirBench Ltd 14 Grange Farm Road Colchester CO2 8JW www airbench com 44 0 1206 791191 Copyright AirBench Ltd 2016 Page of 1 16 AIRBENCH LT...

Page 2: ...ENERAL 7 IN HOUSE COMMISSIONING 7 Operation 8 PRIOR TO USE 8 GENERAL USE 8 Maintenance 9 FILTER CARE 9 SPARES 10 OTHER ROUTINE MAINTENANCE 11 Accessories 12 FILTER PRESSURE GAUGES 12 HOURS RUN METER 1...

Page 3: ...ach unit is marked with its design application which is also shown on the commissioning certi cate The lters are not self cleaning They must be maintained cleaned replaced as described within Do not m...

Page 4: ...ystem which utilises a fan and lter combination selected for your speci c application to extract and lter dust fumes and return ltered air to the work place Some models vary in ltration and ltered air...

Page 5: ...position the top section on the base section always ensure operators are fully trained and that other users are outside the working zone of the forklift Once loosely positioned ensure mounting holes a...

Page 6: ...ceiver of 5l volume within this unit Ducted Units As standard VertExes are con gured to discharge upwards through the top of the unit without ductwork Ducting should not be retro tted to the discharge...

Page 7: ...the lter maintenance routine required for this machine Once complete you should note this on the commissioning certi cate and retain the certi cate for future reference As operating conditions vary e...

Page 8: ...gauge is showing in the upper red zone check the lter condition they may require cleaning or replacement This can also be checked against the lter inspection pressure drop readings in the Commissioni...

Page 9: ...nying log sheet MAIN FILTER ACCESS Main Cartridge Filters are accessed by turning the quarter turn latches on the front cover to loosen and opening the front panel Filters can be released by undoing t...

Page 10: ...See the speci c list for your machine shown on the commissioning certi cate When ordering please contact the manufacturer quoting the serial number of this machine Copyright AirBench Ltd 2016 Page of...

Page 11: ...f moisture If moisture is present within the tank stop and have it examined by a specialist FILTER SEALS These sit behind the lters and prevent dust passing into the body of the bench and the exhaust...

Page 12: ...switch is powered Wheels Standard industrial rubber wheeled swivel castors are used The two front wheels are lockable Lock by pressing on the lever and trip by icking models and operation vary slight...

Page 13: ...ed and recorded within the system logbook Condition check of fan lters electrics Performance testing according to recommended face velocities Visual check that operators are working within effective a...

Page 14: ...TECHNICAL Arrangement 1274 700 2465 1392 Copyright AirBench Ltd 2016 Page of 14 16...

Page 15: ...nal HEPA outlet lter Check the Commissioning Certi cate for details speci c to your unit Model Width Height Depth Nominal Vent Height Weight Fan Power Max Current mm mm mm mm kg kw a VP121218 1270 246...

Page 16: ...achines general requirements EN 60335 2 80 Safety requirements for electric fans and regulators We AIRBENCH Limited declare that VertExes when supplied as self contained equipment comply with the dire...

Reviews: