Other routine maintenance
We recommend the following tasks be performed on a regular basis. Suggested intervals are shown and these
should be reviewed dependent on level of use. Actual completion of these operations and any variation to
these recommended maintenance intervals should be recorded in the system logbook.
If completion of routine maintenance tasks suggests additional issues, contact the manufacturer for advice.
AIR TANK
Check for presence of moisture. If moisture is present within the tank - stop, and have it examined by a
specialist.
FILTER SEALS
These sit behind the filters and prevent dust passing into the body of the bench and the exhaust airstream. It is
important the seals are in good order. Check the condition of the filter seals every time filters are changed or
removed.
VENTILATED DOOR
At least monthly, or as routine, check the surface condition. Clear any blocked holes.
FAN CHAMBER
At least every 6 months, check the fan chamber. Remove any debris and clean.
FAN
At least annually, check and tighten fan fixings, impeller fixings, and electrical connections.