Air Techniques, Inc.

Page 18

Configure the Intraoral Plate 
Guides. 

If the desired Plate Guides 

are not in place, install these Guides 
at this time . Do not operate the scanner 
without a full complement of two Guides 
in place .  Any combination of Guides 
may be used . See Figure 9 .

 

Figure 9 . ScanX with Plate Guides in Place .

ImaGInG pRocedURes

Scanner Track 2 

Plate Guide

Scanner Track 1 

Plate Guide

IMPORTANT:

Transfer the PSP from the PSP 

Transfer box to the Plate Guide slot 

quickly. Always keep the sensitive side 

away from any light source to minimize 

image loss.

Figure 10 .

PSP Pick-up .

Figure 11

Feeding Intraoral Plate .

Activate Scanner

Activate ScanX by performing the following procedures .

1 .  Make sure ScanX and computer components are correctly connected as shown in Figure 4 .
2 .  Switch the scanner from Standby to ON by pressing the membrane POWER switch (    ) located on 

the Keypad Control Panel on the top of the scanner .

3 .  Verify that both the green and blue LED indicators above the POWER and ERASER switches, 

respectively, illuminate . (Default mode has eraser ON .)

4 .  Use the user-supplied authorized Imaging Software to activate the Scanner and to select the desired 

image type and resolution . 

5 .  Verify that the two Scanner track indicators illuminate green when the Scanner has been activated, 

indicating that PSPs can be fed into the corresponding Plate Guide .

Scanning and Erasing Plates

Scan and erase an intraoral PSP in one operation by performing the following procedures .

1 .  Grasp a PSP by short edges between your thumb and index finger .  See Figure 10 .
2 .  With printed side facing you, carefully and quickly insert the PSP into the corresponding Plate Guide 

slot as far as possible . 

3 .  Immediately press the PSP all the way down with your fingertip (from position Figure 11 to position 

Figure 12) until the scanner's transport mechanism takes over and the PSP moves down on its own .

4 .  Verify that the track light illuminates yellow, indicating the Plate Guide slot is in use, the PSP has 

been sensed and the Scanner is transporting the PSP . 

Figure 12

Fully Inserted Plate .

Summary of Contents for ScanX Duo

Page 1: ...ScanX Duo Digital Imaging System Instruction Manual...

Page 2: ...s a leading manufacturer of dental medical and veterinary equipment since 1962 The ScanX Duo is designed to process intraoral imaging plates only See the sections Unpacking and Inspection and Technica...

Page 3: ...or symbols are used on the equipment on the serial number label or in this manual to denote information of special importance DANGER AVOID DIRECT EXPOSURE TO BEAM LASER RADIATION WHEN OPEN CAUTION CLA...

Page 4: ...of ScanX is only accessible by removing hardware with tools It should be opened and ser viced only by an authorized dealer service technician Failure to heed this warning may result in equip ment dam...

Page 5: ...a trained technician from an authorized dealer should remove a cover from ScanX Direct eye contact with the output beam from the laser may cause serious damage and possible blindness Do not open ScanX...

Page 6: ...ich accumulate condensation or become wet through a change of temperature may only be operated after they are fully dry Air Techniques regard themselves as being responsible for the equipment with reg...

Page 7: ...e The manufacturer is not to be held liable for any damage caused as a result of incorrect usage The operator bears all risks This manual provides the information necessary for the setup operation and...

Page 8: ...item is missing notify your dealer Unpacking and inspection System Components Description Quantity Part No 10 Foot Power Cord 1 73096 6 Foot 2 0 USB Cable 1 B3554 Accessory Kit containing Pack of 4 S...

Page 9: ...rvice Pack 3 or later for an Intel 32 bit processor Microsoft Windows XP Professional 64 bit Edition with Service Pack 2 or later for an Intel 64 bit extended x64 processor Microsoft Windows Vista Bus...

Page 10: ...indicator Illuminates steady blue to indicate that the Erase function is ON and PSPs will be erased after scanning Flashes blue to indicate that the Erase function is OFF and PSPs will not be erased...

Page 11: ...d Resolution Range Horizontal 10 to 12 lp mm Vertical 10 to 18 lp mm Compliance Data Laser Class Class I Laser Product 21 CFR 1040 10 Class 1 Laser Product IEC 60825 1 Laser Product Report Accession N...

Page 12: ...s AC or a c alternating current CD ROM compact disk read only memory CFR Code of Federal Regulations CPU central processing unit your computer cm centimeter D depth GB gigabyte 230 109 bytes GHz Giga...

Page 13: ...e computer meets all requirements listed on page 9 3 Verify that an authorized Imaging Software is installed properly on the computer 4 Verify that the supplied ScanX Drivers and Utilities Disk which...

Page 14: ...Press HERE to turn ON or OFF Standby Main Power Switch 1 0 See Note IEC Connector 2 A Circuit Breakers TO AC OUTLET Line Cord Keypad Control Panel USB Connector Type B USB TO COMPUTER POWER Switch ER...

Page 15: ...must always be erased prior to use Note Use PSPs within 24 hours of last erasure Repeat erasing process if PSPs have been stored longer than 24 hours Erase the PSP Each Intraoral PSP should be used i...

Page 16: ...Plate surface Wipe back and forth and then in a circular motion 2 Allow the plate surface to air dry Make sure that the PSP is completely dry before re using Disinfecting the Phosphor Storage Plates T...

Page 17: ...lates in the X ray session have been exposed prepare each plate by performing the procedure below Preparing the Exposed Plate for Scanner Processing 1 Disinfect the Barrier Envelope with plate still i...

Page 18: ...th the green and blue LED indicators above the POWER and ERASER switches respectively illuminate Default mode has eraser ON 4 Use the user supplied authorized Imaging Software to activate the Scanner...

Page 19: ...when removing from the scanner outlet 10 View and save each scanned image using the user supplied authorized Imaging Software IMPORTANT PSPs will not be erased after scanning when operating ScanX wit...

Page 20: ...fect the Plate Guide unless contamination is suspected If a Plate Guide has touched a contaminated surface it may be immersed briefly in a cold sterilant according to sterilant manufacturers direction...

Page 21: ...or white side remains relatively smooth This scratched look on the black side has absolutely no effect on the quality of the image and should be expected under normal conditions If the phosphor side...

Page 22: ...m Verify Computer requirements Page 9 Make sure that the POWER switch is set to ON and the green indicator light is lit Reconnect the cable Check for tightness Replace if necessary Verify that the Set...

Page 23: ...ge due to exposure to light during handling of the PSP Make sure ScanX is operating with eraser turned on blue LED indicator above the ERASER switch is illuminated steadily Make sure the plates are in...

Page 24: ...s Size Order No Qty 0 73248 0 100 1 73248 1 100 2 73248 2 300 3 73248 3 100 4 73248 4 50 ScanX Cleaning Sheets Description Order No Qty ScanX DUO Sheets D1010 12 ScanX DUO Sheets D1020 25 One re usabl...

Page 25: ...al imaging ScanX Systems and Polaris Intraoral Cameras Key Features Easy to learn and simple to use Acquire mount view and store images with just a few mouse clicks Customize images to personal prefer...

Page 26: ...replaced at our option at no charge provided that our inspection shall indicate it to have been defective and that the system is returned to our factory properly received Dealer labor shipping and han...

Page 27: ...care professionals only This equipment may cause radio interference or may disrupt the operation of nearly equipment It may be necessary to take mitigation measures such as reorienting or relocating t...

Page 28: ...ine s 2 kV line s to earth 1kV N A Mains power quality should be that of typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply input lines...

Page 29: ...rom fixed RF transmitters as determined by an electromagnetic site survey a should belessthanthecompliance level in each frequency b Interference can occur nearby devices bearing the following symbol...

Page 30: ...er W Separation distance as a function of the transmitter s frequency m 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz d 1 16 P d 1 16 P d 2 33 P 0 01 0 116 0 116 0 233 0 1 0 366 0 366 0 736 1...

Page 31: ...Page 31 Air Techniques Inc Notes...

Page 32: ...hnical support Whether the need is digital imaging utility room equipment or merchandise Air Techniques can provide the solution via our network of authorized professional dealers Proudly designed tes...