Air Liquide Presence PR Instructions For Use Manual Download Page 7

3. Troubleshooting guide

Fault/concern

 

No flow of gas

Cannot  
connect to the 
Outlet

Hissing noise

Possible cause

Cylinder is empty

Outlet nozzle may 

be blocked

Flow selector  

in-between settings

Blocked port in flow 

selector

Hose diameter is 

wrong size

Shroud is twisted

Damaged sleeved 

outlet

Gas leak

Response to be taken

Check that the flow selector knob or any attached device has not been 

left ON. If contents have been depleted then place unit in the empty 

cylinder storage area and replace with full one.

If it’s a first time use, then check the pressure gauge and if the gauge  

is in the red section immediately contact customer service and  

quarantine the unit for investigation by an Air Liquide Healthcare  

representative.

DO NOT attempt to clear a blocked nozzle. Immediately contact  

customer service and quarantine the unit for investigation by an Air 

Liquide Healthcare representative.

View the top of the flow selector knob to ensure the flow setting  

is visible through the viewing port. If the number is not clearly visible 

then rotate the knob till the required flow is clearly visible and clicks 

into place.

Immediately contact customer service and quarantine the unit for  

investigation by an Air Liquide Healthcare representative.

Check that the tubing fits firmly over the barbed outlet. If not then  

replace the tubing.

View the front of the unit and if the shroud is twisted contact  

customer service and quarantine the unit. Obtain another unit for use.

If the unit is required immediately, then twist the shroud back in-line 

and use as normal. At the completion of use, contact customer  

service and quarantine the unit for investigation by an Air Liquide 

Healthcare representative.

Check that the thread is not damaged to either the sleeved outlet or 

the device it’s being connected to.

Check that the outer rim of the sleeved outlet has not been  

damaged and is circular in shape. Check the hand wheel engages 

properly.

If sleeved outlet is damaged then contact customer service and  

quarantine the unit for investigation by an Air Liquide Healthcare  

representative.

First check to make sure that the flow selector knob has not been left ON.

Do not attempt to stop a leak. Turn the valve OFF and immediately 

contact customer service and quarantine the unit for investigation by 

an Air Liquide Healthcare representative.

•  Read the product label and SDS before using.
•  Do not use grease or oil or any oil-based product.
•    Do not use or store near a naked flame, a source of 

ignition or combustible material.

•   Do not smoke nearby and use appropriate signage.
•    Use and store in a well ventilated area and store in a 

secure and upright position.

•    Do not connect any device to the fill port located at the 

rear of the integrated regulator-valve. The metal cap 
shall NOT be removed under any circumstance. Refer 
to the warning label located near the fill port. 

•    Only attach fittings and accessories to the integrated 

regulator-valve that are compatible with medical oxygen

•    Do not store Presence™ units in any area where  

insect infestation may occur. Do not place objects 
inside the nozzles and ensure proper protection 
of the outlet nozzles when not in use by using the  
plastic caps as supplied. 

•    Flow inaccuracies can be caused by back-pressure 

downstream of the integrated regulator-valve.

•    Any repair, service or modification shall only be  

carried out by an authorised agent.

SAFETY INFORMATION

7

Summary of Contents for Presence PR

Page 1: ...www airliquidehealthcare com au PresenceTM PR PRC PRE Instructions for Use...

Page 2: ...ed by Air Liquide Medical Systems 6 rue Georges Besse 92182 Antony France and is supplied as part of the Presence range of products in Australia by Air Liquide Healthcare Pty Ltd 5 476 Gardeners Road...

Page 3: ...outlet or a fixed flow sleeved outlet PresenceTM PresenceTM PresenceTM PRE PR PRC Key features 3 Easy to use and ergonomic for a quick and easy set up 3 Eliminates the cost of additional equipment sep...

Page 4: ...arged and check that the pressure indicated on the gauge matches the pressure on the product label If the cylinder has already been used check the pressure gauge to ensure there is sufficient medical...

Page 5: ...n is flowing from the barbed outlet before administering to patient e When therapy has concluded turn off the flow by turning the flow selector knob in an anti clockwise direction until it no longer m...

Page 6: ...l f When required remove the attached equipment from the sleeved outlet and replace the cap g When attaching a suction unit to a PRE cylinder the use of a small extension piece will be required refer...

Page 7: ...contact customer service and quarantine the unit for investigation by an Air Liquide Healthcare representative Check that the thread is not damaged to either the sleeved outlet or the device it s bei...

Page 8: ...SA 95 Fullarton Road Kent Town SA 5067 Tel 08 8331 1648 eFax 02 9364 7477 VIC TAS 40 Bunnett Street North Sunshine VIC 3020 Tel 03 9300 5644 eFax 02 9364 7482 WA NT 51 Hampden Rd Nedlands 6009 Tel 08...

Reviews: