1.  Inspection checks to be performed by the user 

before initial use

a.

 

     Check that a clear plastic heat shrink wrap with Air 

Liquide Healthcare logo is intact, and fully covers the 
shroud (refer Fig 1). Check that the cylinder is clean and 
free from oil and grease.

b.

 

    Check outlet protective caps are in place prior to  

removing wrap. When possible always replace the caps 
when not in use (refer Fig 1).  

c.

 

    Check that the valve is in the red “O” OFF position (refer 

Fig 2). View the front of the valve assembly and identify 
all the components (the barbed and sleeved outlets and 
the pressure gauge).   

d. 

     Check that the pressure gauge shows the cylinder is 

fully charged and check that the pressure indicated on 
the gauge matches the pressure on the product label. If 
the cylinder has already been used, check the  
pressure gauge to ensure there is sufficient medical 
oxygen left. The coloured sections are an indication of 
the available contents:

        •  Green: Sufficient contents 
        •  Yellow: Contents are getting low 
        •   Red: Contents are low. Discontinue use and  

 replace with a full one  

e.

      

Check the product label to ensure that the gas matches 
the application. Read the operating instructions  
provided on the label fitted to the side of the shroud 
(refer Fig 3).   

If any deviation is observed do not use the cylinder  
and contact your nearest Air Liquide Healthcare  
office on 1300 360 202.

Figure 1 :  Heat shrink wrap 
and protective caps

Figure 2:  Outlets and 
pressure gauge

Figure 3: Operating 
instructions and product 
label

4

a

b

c

d

e

Summary of Contents for Presence PR

Page 1: ...www airliquidehealthcare com au PresenceTM PR PRC PRE Instructions for Use...

Page 2: ...ed by Air Liquide Medical Systems 6 rue Georges Besse 92182 Antony France and is supplied as part of the Presence range of products in Australia by Air Liquide Healthcare Pty Ltd 5 476 Gardeners Road...

Page 3: ...outlet or a fixed flow sleeved outlet PresenceTM PresenceTM PresenceTM PRE PR PRC Key features 3 Easy to use and ergonomic for a quick and easy set up 3 Eliminates the cost of additional equipment sep...

Page 4: ...arged and check that the pressure indicated on the gauge matches the pressure on the product label If the cylinder has already been used check the pressure gauge to ensure there is sufficient medical...

Page 5: ...n is flowing from the barbed outlet before administering to patient e When therapy has concluded turn off the flow by turning the flow selector knob in an anti clockwise direction until it no longer m...

Page 6: ...l f When required remove the attached equipment from the sleeved outlet and replace the cap g When attaching a suction unit to a PRE cylinder the use of a small extension piece will be required refer...

Page 7: ...contact customer service and quarantine the unit for investigation by an Air Liquide Healthcare representative Check that the thread is not damaged to either the sleeved outlet or the device it s bei...

Page 8: ...SA 95 Fullarton Road Kent Town SA 5067 Tel 08 8331 1648 eFax 02 9364 7477 VIC TAS 40 Bunnett Street North Sunshine VIC 3020 Tel 03 9300 5644 eFax 02 9364 7482 WA NT 51 Hampden Rd Nedlands 6009 Tel 08...

Reviews: