This guide does not replace the user manual
Intended for all the persons authorized to adjust settings

Quick guide

It is recommended to carry out this test after each
patient circuit and / or expiratory valve replacement.

Press

Autotests

Close

the patient circuit, 

validate 

to start the testing 

sequence (<2 min).

After the beep and the message,

remove

the cap to

end the tests.

Autotests

Contact

Air Liquide Medical Systems
Parc de Haute Technologie
6, rue Georges Besse 
92 182 Antony Cedex - France

Phone  : +33 (0)1 40 96 66 00
Hotline : 0820 14 63 59

Expiratory valve MONNAL EVA 

Extraction :

press the eject

button

Single use 

valve

Autoclavable

valve

Cleaning protocol

(see user manual)

Assembly instructions

It is recommended to use a bacteriological filter on
the inspiratory output, to limit the risk of retro-
contamination (see user manual).

Bacteriological filter

Assembly : 

Integrate

the silicone disks 

into the membrane 

(1)

• 

Reposition 

the membrane in 

the valve body 

(2)

• 

Integrate

the flow sensor 

(3) 

at the valve body 

(2)

(1)

(2)

(3)

(3)

www.device.airliquidehealthcare.com

Y

N

03

35

52

 -

Ju

in 

20

18

 ver
sio

n

Note:

pressing the 

padlock

locks the 

touchscreen. 
To unlock = 

press

the padlock and 

confirm

with the rotary knob.