Page 47
17 Default Settings
As supplied, all conditions are set to defaults as in the listing below.
The meter is returned to default settings by the system reset command from the Utilities menu (see
section 12.8) or by the
RST remote command.
Defaults:
Measurement parameter set to DC Volts, single measurement mode.
Auto-ranging mode set.
All first level modifiers cancelled, including Ohms Null.
All second level modifiers cancelled.
Ax + b factors set to A = 1
.
0000 and b = 00000
Limits set to 0.0000
Delta % reference set to 10000
dB reference impedance set to 600
:
Watts reference impedance set to 50
:
Data logger set to Off and timer set to 0000 seconds (logger store contents preserved)
External Trigger set to Off
Speed set to slow (4 readings/sec)
Measurement update symbol (star) set to On
Beep sound set to Off
Measurement filter set to On
RTD probe type set to PT100
RTD measurement mode set to 4 wire
Bus address set to 1 (not by
RST)
Baud rate 9600 (fixed)
N. B. Calibration constants are unaffected by a system reset.