Doc No. FM0460 issue K Page 6  

Registering a Temperature Extreme Sensor onto a

 

Touchsafe Pro 

Nursecall system

 

 

1. Put the Touchsafe Pro Master Panel into Telecare registration mode; touch the screen and Logon as an Engineer or a Manager 

(password required), 

then from the main menu page touch the 

Telecare

 

button to open the screen shown below…

 

 
 

 

 

 

 

 

 

 
 
 
 

 

 

 

 

 

 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

2. Enter the temperature extreme sensor 8 digit 

ID’s…

 

   

 

 
 
 

 

 

 

 

 

 

 

 

 
 

 

 
3. Press the 

Close

 button to exit Telecare mode and the Nursecall panel is now back in normal operating mode. 

 

4. Press the TEST button and check a high temperature alarm call is displayed on the Nursecall panel like below followed by a low 
temperature alarm call about 20 seconds later. 

Remember to remove and refit the battery after the test call.

  

 
 
 
 
 
 
 
 
 

 

The screen banner is 

red

 for high/low temp alarm calls (

Touch the banner to display notes

). 

A high or low temperature  alarm call cannot be accepted or silenced from the screen.  
You must visit the appropriate Nursecall Call Point to cancel the call. 

 
Deleting a Temperature Extreme Sensor from a

 

Touchsafe Pro 

Nursecall system

 

 

Select the high temperature sensor ID to be deleted from the 

Delete Telecare Device

 pull-

down list box (prefixed “T4

1

”).

 

Details of the sensor selected will be displayed in the 

Room

House Code

 and 

ID

 boxes below. 

If the details shown are correct press the 

Delete

 

button.  

Select the low temperature sensor ID to be deleted from the 

Delete Telecare Device

 pull-

down list box (prefixed “T4

2

”) 

 

Repeat as above and press the

 

Close

 

button to exit. 

 

 

 
 

Refer to the Touchsafe Pro Display Panel user guide (Doc No. FM5505) provided with the Nursecall system for 
more information.  

 
 

 

Tynetec and Aidcall operate a policy of continuous improvement and reserve the right to change product specifications without notice. 
If any variation to 

the details contained in this document are suspected please contact Tynetec’s customer support dept on 01670 352371. 

 

The 8 digit ID

’s are mad

e up of the 2 digit device type code (High Temp = 41 & Low Temp = 42)   

followed by the 6 digit unit ID which can be found on a label on the back of the sensor. 

 

 

Touch the 

Add Telecare Device

 box then use the on-screen keypad to enter the high temp ID.  

 

Hint: enter a zone in the 

Filter

 box to narrow the search for the room (see below).  

 

Select the users Call Point ID from the 

Room

 pull-down list box then press 

Save

Note:  this is the room unit that will be will be used to cancel the call. 

 

Touch the 

Add Telecare Device

 box again and repeat as above but this time enter the low 

temp ID i.e. 42

 

followed by the same 6 digit ID from the label

Use keypad to enter  

8 digit ID

’s*

 here 

Select the relevant  

Call Point ID here 

Save when done 

Close to exit mode 

On-screen keypad 

CALLING 
CANCEL 
HC 345  ID 001

 

Press the 

LIGHT GREY

  

button to cancel the call 

*Note: both high and low 

temperature ID

’s 

must 

be entered; 

  Prefix 41 = High Temp 

and Prefix 42 = Low Temp 

         

ZXT426

    W66265 

Ser: 10151997 
ID: 037233 
EE1770 V1.05A 

DR 

 

 