Main specifications 

 

Instantaneous / Integrated Flow Rate Indicator 

Alarm Instantaneous Flow Rate 

Indicator 

Model NH

□□□

-4 NK

□□□-

Display and 

digits 

LCD, 8 digits 

LCD, 4 digits 

Display unit 

L, m

3

 and L/min, m

3

/min 

L/min and m

3

/min 

Output signal 

* Open drain output (equivalent to open collector) 
Pulse width: 

5 ms or more, 

Maximum rated voltage:  30VDC 
Output capacity:    ON  resistance:  150

 or less, 

OFF resistance: 100k

 or more 

(Remaining voltage is 1.5V or less when input 
current is 10 mA or less) 

N/A 

Alarm output 

and setting 

N/A 

Non-voltage c contact output 

110VAC 0.3A, 20VDC1A or less

High / Low setting for alarm -- 

one for each. 

2-digit digital setting 

Other functions 

6-digit trip integration, 

Constant setting 

Integrated value zero clear, 

Hold 

Constant setting, Hold 

Weight 

Approx. 210g 

Approx. 250g 

 
NH / NK common specifications 

 Input signal 

Open collector input 
(supporting 4-line type of ND type flow rate sensor and OF-Z flow rate sensor) 

 Power 

Built-in lithium battery. Battery life is about 10 years under normal use. 
* No battery replacement. 

 Installation method 

Panel embedded type 

 Operating temperature range 

0 to + 60

°

 Operating humidity limit: 30 to 90 %RH (No dew condensation) 

 Receivable distance 

30m 

Number of the input transmission lines

 

 
* For 



 in the table above, connectable meters are as follows. 

005: ND05 

010: ND10 

020: ND20 

05M: OF05Z 

10M: OF10Z 

* The pulse unit of output signal is specified on the nameplate. 

Standard specifications are as follows. 

NH005-4: 10mL/P 

NH010-4: 1L/P 

NH020-4: 1L/P 

NH05M-4: 10mL/P 

NH10M-4: 10mL/P 

 

 
 
Common to NH and NK 

 Connect the wires with the meter correctly in accordance with the connection diagram. 

 Keep the wiring distance from the meter within 30 m. 

 Keep the distance of 20 cm or more from the power line (100VAC, 200VAC) for connection with the meter or this 

Indicator or between them. 

 If the Indicator is affected by noise from a relay, motor, etc., install a spark killer or otherwise for noise prevention 

at the noise source. 

 Be careful that this product may malfunction near equipment that generates harmonics, such as mobile phone, 

ultrasonic cleaner, high frequency generator, or transceiver. 

 
NH 

 When using the output, pay attention to the output polarity. Keep the output line at least 20 cm away from the 

power line 

 Shielded cable is recommended for wiring using the output. 

The shielded cable must be grounded on the side of control equipment. 
When not using the output pulse, the shielded cable must be grounded on the side of this Indicator. 

 
 

 

 
 
NK 

 Do not exceed the rated value when using contact point. 

 
 
 
 
Factory default values 

Mode 

Mode 1 

Mode 2 

Mode 3 

Mode 4 

Mode 5 

Mode 6 

Mode 7 

NH005-4 

2500 6  02 02 02 2.0  1 

NH010-4 

7690 6  01 01 01 2.0  2 

NH020-4 

2500 5  01 01 01 2.0  2 

NH05M-4 

4600 7  03 03 03 2.0  2 

NH10M-4 

2500 6  02 02 02 2.0  1 

* For the factory default values of NK, replace NH above with NK. 

For NK, setting items are from Mode 1 to Mode 5. 

 

Flow Rate Indicator NH / NK

 

I n s t r u c t i o n   M a n u a l

 

Precautions in handling

 

Meter 

Controlled 
equipment 

D type

2 cores

 

4 cores

NH

Meter

4 cores

NK 

Summary of Contents for NH***-4 Series

Page 1: ...ndard specifications are as follows NH005 4 10mL P NH010 4 1L P NH020 4 1L P NH05M 4 10mL P NH10M 4 10mL P Common to NH and NK Connect the wires with the meter correctly in accordance with the connect...

Page 2: ...splay or CL display is blinking the screen will be switched to the integrated flow rate display In switching the integrated flow rate value will not be reset If pressed simultaneously for 2 to 4 secon...

Page 3: ...display digit on the right will blink 4 Each time the L min switch is pressed the setting value will change in turn Setting value is 1 0 1 2 3 4 or 5 5 When the value is set with the TOTAL TRIP switc...

Page 4: ...the third digit of the integrated flow rate display is counted upward a pulse is outputted 3 When the L min switch is pressed for 2 seconds the set value will be lit and setting will be completed Not...

Page 5: ...equivalent value of pulse constant If the SET HOLD switch is pressed when Mode No 1 is lit the screen will switch to Mode No 2 and display the following shows the present setting value 1 When the SET...

Page 6: ...sec 3 When the SET HOLD switch is pressed for 2 seconds the set value will be lit and setting will be completed When the SET HOLD switch is pressed in this state the screen will return to Mode No 1 N...

Page 7: ...tructive act such as a crime 3 Caused by corrosion due to use in a corrosive environment 4 Caused by acts of animals such as a dog cat rat or insect 5 Caused by a factor other than our product 6 Which...

Reviews: