Operating Instructions cont...
Check baby’s positioning before closing Baby Pod 20. Re-check the security of the Infant Positioning Straps before
closing the device.
Carefully close Viewing Lid back over the Baby Pod 20 Outer Shell, using two hands at all times until the Lid
locates against the sides of the Outer Shell. Take care at all times that any ventilator or IV tubing passing through
the Tubing Access Ports on the sides of the Baby Pod 20 do not become trapped under the Viewing Lid. Gently
slide the front lid section forwards towards the Head of Baby Pod 20, and relocate on the lid locking mechanism,
checking that the mechanism is securely closed and that the lid cannot slide backwards of its own accord.
DO NOT
transport a baby without verifying the security of the Viewing Lid.
Re-check the security of the Stretcher Fixing Straps before transporting baby. If straps have become loose, tighten
by grasping the webbing and pulling away from the slider buckles.
Re-check the security of Baby Pod 20’s closure to ensure patient safety.
Section 6.8 - Monitoring baby inside Baby Pod 20
Monitor baby visually while using Baby Pod 20.
Check baby’s skin occasionally for redness or irritation while TransWarmer Infant Transport Mattress is in use.
Monitor infant’s temperature while TransWarmer Infant Transport Mattress is in use.
To minimise heat loss during transport, leave patient access port in Viewing Lid in the closed position during
transport unless the baby requires active care.
Section 6.9- Removing baby from Baby Pod 20
To remove baby from Baby Pod 20, release the black lid lock mechanism at the Head end of Baby Pod 20 and
gently slide the front section of the viewing lid towards the back of the device. When the lid is fully retracted and
cannot slide any further, carefully lift the entire viewing area upwards and over the back of the device, using two
hands at all times, until it rests on the surface on which the device is placed, being careful not to bump the viewing
lid on any surrounding objects.
Remove Infant Positioning Strap from across the baby’s torso, being careful not to disturb any ventilator or IV
tubing.
Open the Vacuum Mattress vacuum valve by pressing the two valve sections together. The click indicates proper
engagement. The mattress should fill with air and become soft and moldable.
Carefully lift the baby from Baby Pod 20, making sure that any ventilator or IV tubing clears the patient
compartment.
Section 6.7 - Closing Baby Pod 20 with baby enclosed
Warning
Patients may be harmed by the incorrect use of this device.
Warning