  10. Device configuration 

10.3 Function locking 

Since there is a risk that the measuring functions or relative measuring opera-

tions may lead to misinterpretation, these are normally locked at level 3. These 

can, however, be individually enabled via menu item ´

Loc

´ : 

Loc

 parameter

 Locked functions 

        0 

None 

         1

Measuring functions (the previously selected function is retained) 

         2

Relative measuring 

         3

Measuring functions and relative measuring (default) 

11. TROUBLE-SHOOTING 

The THERM 2420 measuring instrument can be configured. If in certain cir-

cumstances the device does not respond exactly as it should, the cause of 

such unexpected behavior is only very rarely a device defect; more usually it is 

incorrect operation by the user, an invalid setting, or unsuitable cabling. In such 

event try to pinpoint and clear the problem with the aid of the following tests. 

Error: 

No display, display malfunction, keys do not react 

Remedy:

 Check the power supply, charge the battery, switch off and then on 

again. If necessary, re-initialize; (see 6.2)

Error: 

Measuring functions or relative measuring cannot be started. 

Remedy: 

Check device configuration; see 10, 10.3

Error: 

Measured values are incorrect. 

Remedy: 

Switch Device OFF / ON, press key and hold

 

 CLR 

Check measur-

ing range configuration; see 10

Error: 

Fluctuating measured values or the system hangs in mid-operation. 

Remedy:  

Check   connecting   cable(s),   connect   hand-held   sensors   in   air   or 

phantoms (for thermocouples short-circuit) and check. 

Then reconnect the sensor and check; if necessary, insulate the sen-

sor and eliminate interference by using shielded or twisted wiring. 

Error: 

´

CALEr

´

 is displayed when the device is switched on. 

Remedy: 

The calibration of a measuring range may have become misadjust-

ed. The device must be recalibrated at the factory. 

If, after performing the above-listed checks and remedial steps, the device still 

fails to behave as described in the operating instructions, it must be returned to 

our factory in Holzkirchen, accompanied by an explanatory note and error de-

scription. 

 12                     THERM 2420 

Summary of Contents for THERM 2420

Page 1: ...____________________________ Operating instructions Temperature measuring instrument THERM 2420 V1 2 27 11 2007 www ahlborn com...

Page 2: ...V REL Relative measuring 3 Operating keys ON OFF Switch the device on To switch device OFF press and hold down M M Meas point selection MAX MIN Max min value clear press and hold down MEM Measured va...

Page 3: ...ation 6 7 CONNECTING THE TRANSDUCERS 6 8 DISPLAY AND KEYPAD 7 8 1 Display 7 8 2 Keypad 8 9 MEASURING OPERATIONS 9 9 1 Measured value 9 9 2 Measured value functions 9 9 3 Peak value memory 10 9 4 Measu...

Page 4: ...r pos sible use the original packaging material for dispatch and enclose a clear and informative description of the fault and of the conditions in which it occurs This guarantee will not apply in the...

Page 5: ...se substantial measuring errors You are advised therefore to wait until the device has adjusted to the ambient temperature before starting to use it Do not run sensor lines in the vicinity of high vol...

Page 6: ...ry capacity drops to approx 10 percent the LOBAT arrow will appear in the display If the batteries are completely dis charged the device will switch off To replace old batteries first unscrew the batt...

Page 7: ...omprises 41 2x 7 segment digits b for depicting various measuring functions a there are also 2 arrows e for depicting the operating status Function field Main field Display of measuring functions in t...

Page 8: ...nction is locked This appears briefly Overshoots measuring range Maximum value flashes Undershoots measuring range Minimum value flashes 8 2 Keypad To operate the device a keypad with 7 keys is provid...

Page 9: ...on field you will see the abbreviation for the thermocouple see Section 10 1 All special operating states possible for the measured value are explained in Section 8 1 9 2 Measured value functions By p...

Page 10: ...20 any measured value can be saved To save the measured value press key MEM The currently saved value then ap pears in the function field preceded by the symbol M To jump back to the operating range d...

Page 11: ...fy the value press keys M or M Entry is completed by again pressing key ON To terminate configuration at any time press key MEM 10 1 Measuring ranges The following measuring ranges can be set via menu...

Page 12: ...eck the power supply charge the battery switch off and then on again If necessary re initialize see 6 2 Error Measuring functions or relative measuring cannot be started Remedy Check device configurat...

Page 13: ...2kV IEC 61000 4 3 1995 A1 1998 A2 2000 3V m The following advisory notes must be observed when operating the device 1 If the standard sensor is extended 1 5 meters care must be taken to ensure that t...

Page 14: ...ype J 200 0 950 0 C Cu CuNi Type U 200 0 600 0 C Cu CuNi Type T 200 0 400 0 C PtRh10 Pt Type S 0 0 1760 0 C Millivolt 10 000 55 000 mV Standard equipment LCD Measured value 5x 7 segment 15 mm 2x 16 se...

Page 15: ...maximum value 9 3 10 Measured value 9 1 9 Measured value functions 9 2 9 Measured value memory 9 4 10 Measuring input 13 1 14 Measuring operations 9 9 Measuring ranges 10 1 11 Order no 13 2 14 Peak va...

Page 16: ...13 Appendix 13 4 Your contact 16 THERM 2420...

Reviews: