5. DATA ACQUISITION 

The instrument ALMEMO

®

 8990-8 provides the following options for the

acquisition of measuring data:
1. Continuous measurement of a selectable measuring point, see manual 6.4
    Possibly, an output of measuring data to an analogue output, see man. 5.1.1
2. Single measuring point scan, see manual 6.5.1.1.
3. Cyclic measuring point scan, see manual 6.5.1.2.
4. Continuous measuring point scan, see manual 6.5.1.3.
Previous measuring data should be cleared before a measurement. 

total clearing of all measured values

, i.e. the clearing of max, min and

average values of all channels, as well as the clearing of the memory is
performed by pressing the keys STOP and OUTPUT. For automatic clearing
during start, see manual 6.10.13.2.

The programming of the process control 

is performed via serial interface

and by using the configuration software AMR-Control or a terminal (PC) (see
manual section 6). If during programming a programming connector (ZA
8990-PS) is connected to the socket A2 all parameters will be transmitted to
the connector. If such a connector is also connected to socket A2 during
switch-on, the data will be reloaded and the instrument will be programmed
correspondingly. By using several connectors the configuration can be easily
changed. If a programming via interface is not required the connectors can be
ordered factory-programmed to customer specifications.

5.1  Single Measuring Point Scan

If no measuring point scan has been set, only the measured value of the
selected measuring point, beginning with measuring point 0, will be
continuously acquired with the selected conversion rate (see manual 6.5.4).
This is the best possible operating mode for a data logging with analogue
output.
Measuring point scans can be used to acquire, indicate and document data
from the selected measuring point and also from other measuring points. As
long as no print cycle and no measuring cycle are programmed (e.g. after a
reinitialisation, see 3.5) only single measuring point scans are triggered by
using the key START. The control lamp ´CYCLE´ is only illuminated during the
data query and will then turn off. If the time was cleared, it will be restarted, a
measuring point scan of all connected active sensors will be performed and the
measured values will, for one time, be provided as output to a connected
printer. If the memory is activated (see manual 6.5.3) the control lamp
´MEMORY´ will be on during the data query. With each press of the key the
measured values are equally processed with the corresponding measuring
time. If true time has to be indicated, it must first be set (see manual 6.2.8).

   14

ALMEMO

®

 8990-8  

   Data Acquisition  

Summary of Contents for Almemo 8990-8

Page 1: ...Operating Instructions Data Logger ALMEMO 8990 8 V1 3 03 02 2004...

Page 2: ...ernal Power Supply 9 3 4 Battery Buffer 9 3 5 Switch On Off Reinitialisation 10 4 Connection of the Transducers 10 4 1 Transducers 10 4 2 Measuring Inputs and Additional Channels 11 5 Data Acquisition...

Page 3: ...5 1 The interface module RS232 fiber optics Centronics manual section 5 2 The entire ALMEMO networking system manual section 5 3 All functions and their control via the interface manual section 6 A c...

Page 4: ...n measuring junctions can be programmed as function channels and can be processed and printed like normal measuring junctions Furthermore function channels for special measuring tasks are provided to...

Page 5: ...Measured Value A continuous presentation of measuring data from the selected measuring point is provided and also includes automatic zero point correction and optional correction of the measured value...

Page 6: ...y during data storage Measuring Cycle The measuring cycle also programmable between 1s and 59h 59min 59s is for a cyclic scanning with a display of all measured values limit value monitoring including...

Page 7: ...dressed and can be easily networked by a simple connection with network cables or network junctions for longer distances Software The AMR Control software which allows for the entire programming of th...

Page 8: ...charged without pow supp ON off battery operation active 2 CONNECTOR SOCKET U DC mains adapter ZA 5090 NA2 12V 800mA connector cable ZB 5090 EK 7 13V DC connector cable ZB 3090 UK 10 30V DC electr iso...

Page 9: ...ODING SWITCH setting of the device address module number 0 to 9 external access 2nd coding switch for decades internal on the board 6 FUNCTION KEYS START start automatic measuring point scan STOP and...

Page 10: ...START see 5 1 Enter measuring cycle for cyclic storing see manual 6 5 3 Enter time and date of start and end of measurement if required see man 6 6 2 Start automatic storing with the key START see 5...

Page 11: ...the on off switch 1 is continuously illuminated If the green lamp flashes the battery is completely recharged and the charge circuit is switched to trickle charging As a result the power supply can d...

Page 12: ...n be connected to the ALMEMO input sockets 0 to 8 of the measuring instrument 3 For connecting existing sensors it is only necessary to connect a corresponding ALMEMO connector 4 1 Transducers A detai...

Page 13: ...If required the sensor can also be programmed with several ranges or scalings or depending on the pin assignment 2 or 3 sensors can be combined in one connector e g TE Ntc mV V mA V etc The additional...

Page 14: ...e configuration can be easily changed If a programming via interface is not required the connectors can be ordered factory programmed to customer specifications 5 1 Single Measuring Point Scan If no m...

Page 15: ...ected see manual 5 2 the output channel U must be set manual 6 2 2 and an output format List Column or Spreadsheet must be selected manual 6 5 5 output logs see manual 6 6 1 For starting the memory ou...

Page 16: ...er the first scan Within the set cycle the instrument will switch on perform a measuring point scan the control lamps CYCLE and MEMORY will be illuminated for a short time and the instrument will then...

Page 17: ...for inadmissible electrical connection disconnect all suspicious sensors hold hand held sensors in air or connect dummies and check short circuit AB at thermocouples 100 at Pt100 sensors then reconnec...

Page 18: ...factory in Holzkirchen Germany including a short report and possibly control printouts The software AMR Control allows to print the monitor pages including the programming and also to save the termina...

Page 19: ...ousing metal housing H63 x W144 x D219mm 3HU 6DU Operating temperature 10 60 C Storage temperature 30 60 C Humidity of ambient air 10 90 rH non condensing Extent of the Delivery Measuring Instrument A...

Page 20: ...20 ALMEMO 8990 8 Appendix...