6.4 Limit Values
Two limit values (MAX and MIN) can be programmed for each measuring
channel. The exceeding of the limit values is handled as a fault, similar to the
exceeding of the measuring range limits and sensor breakage. The arrow
ALARM will appear in the display and the alarm relays will respond (see
manual 6.3.9). An exceeding can also be used to start or stop a measuring
point scan (see manual 6.6.3).
Function LIMIT VALUE HIGH
´
LH
´ and
LIMIT VALUE LOW
´
LL
´
Selection with key:
...
FUNCTION
1: 1 2 3.0
LH
Programming:
Input according to 5.4
Switch-off:
,
ENTER
Clr
1: - - - -
LH
6.5 Correction Values
The correction values ZERO POINT and SLOPE allow for correcting sensors
with regard to zero point and slope (gain) (see manual 6.3.10).
Corrected Meas. Value
= (Meas. Value - ZERO POINT) x SLOPE.
Function ZERO POINT CORRECTION
´
ZC
´
Selection with key:
...
FUNCTION
1: 0 0 3.2
ZC
Programming:
Input according to 5.4
Clear:
,
ENTER
Clr
1: - - - -
ZC
Function
SLOPE CORRECTION (GAIN):
´
SC
´
Selection with key:
...
FUNCTION
1:1.5 0 0 0
SC
Programming:
Input according to 5.4
If correction values are programmed and, as a result, the measured value is
corrected, the arrow
CORR
will be indicated in the display.