After a reinitialisation (see 3.4) the basic functions (bolded in table) for
operating the instrument with built-in printer are available by using the keys
MEASURE
and
FUNCTION
. More functions are available and can automatically or
manually be activated, as required, for correcting measured values and for
data transfer. This provides fast access and a reduced risk of incorrect entries.
However, to have any function available in case it is required, all functions can
be temporarily activated, i.e. until the instrument is switched off again:
Activate all functions
by switching on with the key
pressed and held.
ENTER
The functions can be activated by repeatedly operating the keys
MEASURE
or
FUNCTION
. If a key is pressed and held for longer than one second the previous
function can be selected again. The functions are identified by a two digit
abbreviation instead of the dimension. In the display the abbreviation for the
function follows the value of the function. In the case of sensor parameters the
channel number is also included:
MEAS. VALUE
FUNCTION
Meas. Functions
Abbrev.
ProgrammingFunctions
Abbrev.
Meas. value
1: 114.3
°C
Output mode
PL 2 OM
Max value
(Hi)
1: 123.4 MH
Paper feed
320.00 PF
Min value
(Lo)
1: 023.4 ML
or
print cycle
00:15:00 PC
Memory (option S)
1: 113.2 MY
Time
12:34:56 TM
&
Date
31.12.99 DA
FUNCTION
Range
1: NiCr R
Memory functions
Plot range - start
1: 000.0 PS
Pplot range - start
1: 000.0 PS
Plot range - end
1: 100.0 PE
Plot range - end
1: 100.0 PE
Limit value Max (Hi)
1: 123.0 LH
Free memory
126.1 FR
Limit value Min (Lo)
1:-010.0 LL
Output mode
PL 2 OM
Locking mode
1: 0005 LM
Start time
10:00:00 ST
Base value
1: ---- BA
End time
10:30:00 ET
Factor
1: ---- FA
Start date
01.05.99 SD
Exponent
1: 0 EX
End date
01.05.99 ED
" Zero point correction
1: ---- ZC
" Slope (gain) correction
1: ---- SC
^ Ambient
temperature
1: 250.0 AT
^ Emissivity
factor
1: 0.950 EF
* Baud
rate
9600 BR
* Device
address
00 A
~
Atmosph. pressure
1013 mb
Automatic activation of the functions by:
*
interface modules
^ infrared sensors
" pH probes
~
humidity, dynam. press. and O
2
sensors
For a data output starting from the measuring function ´
MY
´ the key
FUNCTION
can be used to call various
memory functions
, which are entirely independent
ALMEMO
6290-7B
18