12. Programming using the programming menus
vated. In the
Mode
function program the variant
Monitor
:
Mode : Moni-
tor
Fail-safe mode
The "fail-safe mode" is suitable when scanning is purely software-based; it
merely ensures, in the event of computer failure, that scanning will continue on
an internal cyclic basis. In this mode the cycle programmed in the device must
be longer than that needed for software scanning. Software scanning keeps
resetting the internal cycle with the effect that this cycle is only actually used
as and when software scanning fails; (in Win-Control “safe initialization” must
be switched off).
The internal cycle is started as and when the Win-Control software starts; it
may also have been started previously. When scanning with the internal cycle
no data is output to the interface. In order to record data the memory must
have been activated.
In the
Mode
function program the variant
Fail-safe
:
Mode : Fail-safe
12.2.6 Memory output
The contents of the internal measured value
memory can be output completely or in excerpts
via the serial interface. With each output any
one of the three known output formats may be
used “list”, “columns”, or “table”.
The option to specify partial ranges is available
as it is possible to set the start and end time of
measurements and also possible to select the
number of corresponding identified measure-
ments.
With
external memory cards
(see
12.2.1) there is only one option avail-
able, namely output in table mode of
all the measured data contained in the file most recently used. For
this purpose use only the
key in the
Memory capacity free
function in the
Memory output
menu or in certain measuring menus.
The most sensible approach is to remove the memory card and copy
the files via a USB card reader directly onto the PC. These can then
be imported either into MS-Excel or into Win-Control (as of V.4.9).
Menu
Memory output
:
Set the
output format
(see 12.1.2):
Output format : List
To
select a numbered measuring operation
:
Number: 12-001
In the
Number
function select the number by pressing
<FIRST>
,
<NEXT>
...,
<LAST>
To
select a time frame
:
Enter the start time in format ´
hh:mm:ss
´ :
Start time :
07:00:00
60 ALMEMO
®
5690-2M
* OUTPUT FROM MEMORY *
±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±±
Memory Internal:
512.0 kB
Memory free:
125.8 kB
Residual output:
12.5 kB
Output form:
Columns
Number:
01-001 A
Time: 12:34:56 Date: 01.01.04
Time interval:
Start time:
07:00:00
Start date:
01.01.04
End time:
17:00:00
End date:
01.01.04
ALL NR F TIME ESC