6.6 Scaling, Decimal Point Setting
For indicating the electrical signal of a sensor as a measured value of a
physical variable it is, in most cases, necessary to set a zero point shift and to
perform a multiplication with a factor. The functions EXPONENT, BASE and
FACTOR are available for this. A detailed description of the scaling, including
an example, can be found in the manual section 6.3.11.
Indicated value
= (corrected meas. value - BASE) x FACTOR.
Decimal Point Setting
The FACTOR can be programmed in the range from -2.0000 to +2.0000. For
factors higher than 2.0 or lower than 0.2 a corresponding decimal point setting
must be considered by entering the EXPONENT. The function EXPONENT
´
EX
´ allows for shifting the decimal point as far to the left (-) or right (+) as it can
be indicated on the display. An exponential presentation of the measured
values is not possible.
Function BASE VALUE
´
BA
´
Selection with key
F2
...
...
Input see 5.4
F2
0
7
.
0
0
B
A
Function FACTOR
´
FA
´
Selection with key
F2...
...
Input see 5.4
-
F2
0
.
1
6
3
F
A
Function EXPONENT
´
EX
´
Selection with key
F2...
...
Input see 5.4
F2
- 2
E
X
ALMEMO
4490-2
23
Sensor Programming