![Ahlborn ALMEMO 4490-2 Operating Instructions Manual Download Page 18](http://html1.mh-extra.com/html/ahlborn/almemo-4490-2/almemo-4490-2_operating-instructions-manual_2870478018.webp)
The keys
▲
and
▼
can be used to select all possible ranges in the sequence
given below. When the desired range is indicated in the display it can, by
pressing the key
again, be transferred including the correct dimension to the
▲
connector and the programming can be finished. All other programming values
of the channel will be cleared by this.
Function RANGE
´
R
´
Selection with key
F2...
...
F2
N
i
C
r
R
Example :
range NiCr-Ni
Change meas. range:
,
... or
... ,
ENTER
▲
▼
▲
Transducer
Conn. / Cable /
Sensor
Meas. Range
Dim Display
Pt100-1
ZA 9000-FS
-200.0... +850.0
°C
P104
Pt100-2
ZA 9000-FS
-200.00...+200.00
°C
P204
Ni100
ZA 9000-FS
-60.0... +240.0
°C
N104
NiCr-Ni (K)
ZA 9020-FS
-200.0...+1370.0
°C
NiCr
NiCroSil-NiSil (N)
ZA 9020-FS
-200.0...+1300.0
°C
NiSi
Fe-CuNi (L)
ZA 9000-FS
-200.0... +900.0
°C
FECO
Fe-CuNi (J)
ZA 9000-FS
-200.0...+1000.0
°C
IrCo
Cu-CuNi (U)
ZA 9000-FS
-200.0... +600.0
°C
CUCO
Cu-CuNi (T)
ZA 9000-FS
-200.0... +400.0
°C
CoCo
PtRh10-Pt (S)
ZA 9000-FS
0.0...+1760.0
°C
Pt10
PtRh13-Pt (R)
ZA 9000-FS
0.0...+1760.0
°C
Pt13
PtRh30-PtRh6 (B)
ZA 9000-FS
+400.0...+1800.0
°C
EL18
Au-FeCr
ZA 9000-FS
-270.0... +60.0
°C
AUFE
Ntc type N
ZA 9000-FS
-30.00...+125.00
°C
Ntc
Millivolt 1
ZA 9000-FS
-26.000...+26.000
mV
U 26
Millivolt
ZA 9000-FS
-10.000...+55.000
mV
U 55
Millivolt 2
ZA 9000-FS
-260.00...+260.00
mV
U260
Volt
ZA 9000-FS
-2.6000...+2.6000
V
U2.60
Differential Millivolt 1
ZA 9050-FS
-26.000...+26.000
mV
d 26
Differential Millivolt
ZA 9050-FS
-10.000...+55.000
mV
d 55
Differential Millivolt 2
ZA 9050-FS
-260.00...+260.00
mV
d260
Differential Volt
ZA 9050-FS
-2.6000...+2.6000
V
d2.60
Sensor voltage
ZA 9000-FS
0.00...20.00
V
UbAt
Milliampere
ZA 9601-FS
-32.000...+32.000
mA
I032
Percent (4-20mA)
ZA 9000-FS
0.00... 100.00
%
P420
Ohm
ZA 9000-FS
0.00... 400.00
Ω
Ohn
Frequency
ZA 9909-AK
0... 25000
Hz
FrEq
Pulses
ZA 9909-AK
0... 65000
PULS
18
ALMEMO
4490-2
Sensor Programming